SPIRIT 3000 DENTAL CHAIR SP30 3003

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2015-09-14 for SPIRIT 3000 DENTAL CHAIR SP30 3003 manufactured by Pelton & Crane.

Event Text Entries

[26136912] Upon visual inspection by the authorized (b)(4) distributor, it was identified that the four mounting bolts that attaches the rear cuspidor bracket to the dental chair were loose which caused the top two mounting bolts to break. The distributor has already disposed of the bolts therefore (b)(4) cannot complete the evaluation. The dental chair is over 10 1/2 years old and is past the expected life of the device.
Patient Sequence No: 1, Text Type: N, H10


[26136913] It was reported that a dental hygienist was performing routine medical treatment to a patient when the rear mounted cuspidor fell off the (b)(4) sp30 dental chair. There were no injuries reported.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1017522-2015-00020
MDR Report Key5075914
Date Received2015-09-14
Date of Report2015-08-17
Date of Event2015-08-17
Date Mfgr Received2015-08-17
Device Manufacturer Date2005-02-21
Date Added to Maude2015-09-14
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationDENTAL HYGIENIST
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. FRANK RAY
Manufacturer Street11727 FRUEHAUF DRIVE
Manufacturer CityCHARLOTTE NC 28273
Manufacturer CountryUS
Manufacturer Postal28273
Manufacturer Phone7045877227
Manufacturer G1PELTON & CRANE
Manufacturer Street11727 FRUEHAUF DRIVE
Manufacturer CityCHARLOTTE NC 28273
Manufacturer CountryUS
Manufacturer Postal Code28273
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameSPIRIT 3000 DENTAL CHAIR
Generic NameDENTAL CHAIR
Product CodeNRU
Date Received2015-09-14
Model NumberSP30
Catalog Number3003
OperatorDENTAL HYGIENIST
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerPELTON & CRANE
Manufacturer Address11727 FRUEHAUF DRIVE CHARLOTTE NC 28273 US 28273


Patients

Patient NumberTreatmentOutcomeDate
10 2015-09-14

© 2026 FDA.report