ABBOTT AXSYM SYSTEM 7A83-01

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06 report with the FDA on 2003-12-23 for ABBOTT AXSYM SYSTEM 7A83-01 manufactured by Abbott Manufacturing, Inc..

Event Text Entries

[350117] A physician questioned an axsym bhcg result of 701 miu/ml on a pt suspected of having either a hydatid mole or ectopic pregnancy. The pt previously tested at 3,541 miu/ml. The pt was redrawn yielding a result of 2,082 miu/ml. The sample in question was retested twice yielding results of 2,498 and 3,013 miu/ml. No impact to pt management was reported.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1628664-2003-00017
MDR Report Key507594
Report Source05,06
Date Received2003-12-23
Date of Report2003-12-17
Date of Event2003-11-09
Date Mfgr Received2003-11-21
Device Manufacturer Date1999-02-01
Date Added to Maude2004-01-26
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location0
Manufacturer ContactDR. DAVID SPINDELL, D-09Y6, AP6C-2
Manufacturer Street100 ABBOTT PARK ROAD
Manufacturer CityABBOTT PARK IL 600646092
Manufacturer CountryUS
Manufacturer Postal600646092
Manufacturer Phone8479351869
Manufacturer G1*
Manufacturer Street*
Manufacturer City*
Manufacturer Country*
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameABBOTT AXSYM SYSTEM
Generic NameAUTOMATED IMMUNOASSAY ANALYZER
Product CodeJHE
Date Received2003-12-23
Model NumberNA
Catalog Number7A83-01
Lot NumberNA
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Implant FlagN
Date Removed*
Device Sequence No1
Device Event Key496541
ManufacturerABBOTT MANUFACTURING, INC.
Manufacturer Address1921 HURD DR. IRVING TX 75038 US
Baseline Brand NameABBOTT AXSYM SYSTEM
Baseline Generic NameAUTOMATED IMMUNOASSAY ANALYZER
Baseline Model NoNA
Baseline Catalog No7A83-01
Baseline IDNA


Patients

Patient NumberTreatmentOutcomeDate
10 2003-12-23