DERMAHOOK 1/2 HOOK 10 PKG/BX 6 HKS/PKG 382805

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2015-09-15 for DERMAHOOK 1/2 HOOK 10 PKG/BX 6 HKS/PKG 382805 manufactured by Teleflex Medical.

Event Text Entries

[26173715] (b)(4). The device history review for the product dermahook 1/2 hook (b)(4), lot number 73e1500506 investigation did not show issues related to the complaint. The device sample has not been returned for investigation at the time of this report. The manufacturer will continue to monitor and trend related events.
Patient Sequence No: 1, Text Type: N, H10


[26173716] Alleged event: the rubber bands are snapping. The patient's condition was reported as fine.
Patient Sequence No: 1, Text Type: D, B5


[37862571] (b)(4). Visual inspection: (b)(4) - all samples passed visual inspection with zero defects. Functional evaluation: (b)(4)- force to stretch (200% and 300%) all samples passed. Results are statistically equivalent to results achieved during design verification testing. Elongation at break tests - all samples passed. (b)(4) - force to stretch tests (200% and 300%). All samples passed. After completion of testing, we conclude that the product is performing as expected and meets all design requirements.
Patient Sequence No: 1, Text Type: N, H10


[37862572] Alleged event: the rubber bands are snapping. The patient's condition was reported as fine.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3003898360-2015-00644
MDR Report Key5079195
Date Received2015-09-15
Date of Report2015-09-08
Date of Event2015-09-04
Date Mfgr Received2015-09-10
Device Manufacturer Date2015-05-27
Date Added to Maude2015-09-15
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactJASMINE BROWN
Manufacturer Street3015 CARRINGTON MILL BLVD
Manufacturer CityMORRISVILLE NC 27560
Manufacturer CountryUS
Manufacturer Postal27560
Manufacturer Phone9193614124
Manufacturer G1TELEFLEX MEDICAL
Manufacturer StreetRANCHO EL DESCANSO
Manufacturer CityTECATE 21478
Manufacturer CountryMX
Manufacturer Postal Code21478
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameDERMAHOOK 1/2 HOOK 10 PKG/BX 6 HKS/PKG
Product CodeGDG
Date Received2015-09-15
Returned To Mfg2015-09-14
Catalog Number382805
Lot Number73E1500506
Device Expiration Date2016-11-05
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerTELEFLEX MEDICAL
Manufacturer AddressRESEARCH TRIANGLE PARK NC


Patients

Patient NumberTreatmentOutcomeDate
10 2015-09-15