DIMENSION RXL MAX WITH HM

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2015-09-15 for DIMENSION RXL MAX WITH HM manufactured by Siemens Healthcare Diagnostics Inc.

Event Text Entries

[26642413] A siemens customer service engineer (cse) specialist was dispatched to the customer site. After evaluating the instrument, the cse replaced the source lamp and aligned it. The cause of the discordant tacrolimus results on four patient samples was related to a malfunction of the source lamp. The instrument is performing according to specifications. No further evaluation of the device is required.
Patient Sequence No: 1, Text Type: N, H10


[26642414] Discordant tacrolimus results were obtained on four patient samples on a dimension rxl max with hm instrument. The discordant results were reported to the physician(s), who questioned them. The samples were repeated on the same instrument, resulting lower for samples (b)(6). The repeat results for samples (b)(6) were not provided by the customer. It is unknown which repeat results were reported to the physician(s). There are no known reports of patient intervention or adverse health consequences due to the discordant tacrolimus results.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1226181-2015-00524
MDR Report Key5079319
Date Received2015-09-15
Date of Report2015-08-21
Date of Event2015-08-17
Date Mfgr Received2015-08-21
Device Manufacturer Date2011-01-27
Date Added to Maude2015-09-15
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag0
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMICHAEL METZ
Manufacturer Street511 BENEDICT AVE
Manufacturer CityTARRYTOWN NY 10591
Manufacturer CountryUS
Manufacturer Postal10591
Manufacturer Phone9145242223
Manufacturer G1SIEMENS HEALTHCARE DIAGNOSTICS INC
Manufacturer Street101 SILVERMINE ROAD
Manufacturer CityBROOKFIELD CT 06804
Manufacturer CountryUS
Manufacturer Postal Code06804
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Sequence Number: 0

Brand NameDIMENSION RXL MAX WITH HM
Generic NameCLINICAL CHEMISTRY ANALYZER
Product CodeMLM
Date Received2015-09-15
Model NumberDIMENSION RXL MAX WITH HM
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrN
Device Sequence No0
Device Event Key0
ManufacturerSIEMENS HEALTHCARE DIAGNOSTICS INC
Manufacturer Address101 SILVERMINE ROAD BROOKFIELD CT 06804 US 06804

Device Sequence Number: 1

Brand NameDIMENSION RXL MAX WITH HM
Generic NameCLINICAL CHEMISTRY ANALYZER
Product CodeJJE
Date Received2015-09-15
Model NumberDIMENSION RXL MAX WITH HM
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerSIEMENS HEALTHCARE DIAGNOSTICS INC
Manufacturer Address101 SILVERMINE ROAD BROOKFIELD CT 06804 US 06804


Patients

Patient NumberTreatmentOutcomeDate
10 2015-09-15

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.