COULTER TQ-PREP WORKSTATION 6605429

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2015-09-18 for COULTER TQ-PREP WORKSTATION 6605429 manufactured by Beckman Coulter.

Event Text Entries

[26207542] The customer observed that the sample tube did not have the correct amount of reagent when the sample was prep on the tq prep instrument. The sample line had aspirated air since the reagent bottles were empty. There was a failure of the sensors. No error message alerted the customer that the reagent bottles were empty since the sensors had failed. The fse replaced both reagent sensors to resolve the issue. (b)(4).
Patient Sequence No: 1, Text Type: N, H10


[26207543] The fse (field service engineer) reported that the reagent sensors (reagent a and reagent b sensors) in a coulter tq-prep workstation failed and did not generate the low reagents messages. The customer performs the daily instrument check as per instructions for use and the customer visually confirms the tubes after preparation, which alerted the customer to the issue. Erroneous results were not generated and there was no change or effect to patient treatment in connection with this event.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1061932-2015-01480
MDR Report Key5088649
Report SourceFOREIGN,HEALTH PROFESSIONAL,U
Date Received2015-09-18
Date of Report2015-09-02
Date of Event2015-09-02
Date Mfgr Received2015-09-02
Device Manufacturer Date2014-07-01
Date Added to Maude2015-09-18
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. MIRANDA HOLLAND
Manufacturer Street11800 SW 147TH AVENUE, 32-S08
Manufacturer CityMIAMI FL 331962031
Manufacturer CountryUS
Manufacturer Postal331962031
Manufacturer Phone3053802031
Manufacturer G1BECKMAN COULTER
Manufacturer Street250 S KRAEMER BLVD
Manufacturer CityBREA CA 92821
Manufacturer CountryUS
Manufacturer Postal Code92821
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameCOULTER TQ-PREP WORKSTATION
Generic NameAUTOMATED PIPETTING, DILUTING AND SPECIMEN PROCESSING WORKSTATIONS FOR FLOW CYTO
Product CodePER
Date Received2015-09-18
Model NumberNA
Catalog Number6605429
Lot NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerBECKMAN COULTER
Manufacturer Address250 S KRAEMER BLVD BREA CA 92821 US 92821


Patients

Patient NumberTreatmentOutcomeDate
10 2015-09-18

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.