Event Text Entries

[26226344] The unit om-501 oxygen conserver and oxygen cylinder were returned to the manufacturer 8/28/2015 and 9/8/2015. The conserver unit was visually inspected and tested under normal conditions. The unit showed no leak or malfunction. All port or barbs were inspected and showed no signs of flash or burn mark penetration. There is no seal washer submitted and the tank had residue on the valve with flash and burn marks.
Patient Sequence No: 1, Text Type: N, H10


[26226345] Patient daughter stated she was changing an empty tank in the back seat of her car (no damage to car). The oxygen conserver and tank experience a flash fire. She experienced burns to her left hand and went to burn center. She has since recovered and reports no other issues.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1062191-2015-00002
MDR Report Key5089561
Report SourceCOMPANY REPRESENTATIVE
Date Received2015-09-18
Date of Report2015-09-18
Date of Event2015-07-25
Date Mfgr Received2015-08-24
Date Added to Maude2015-09-18
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. MICHAEL WOJNOVICH
Manufacturer Street401 LEONARD BLVD. N.
Manufacturer CityLEHIGH ACRES FL 33971
Manufacturer CountryUS
Manufacturer Postal33971
Manufacturer Phone2396436577
Manufacturer G1INOVO, INC
Manufacturer Street401 LEONARD BLVD. N.
Manufacturer CityLEHIGH ACRES FL 33971
Manufacturer CountryUS
Manufacturer Postal Code33971
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCYPRESS
Generic NameCONSERVER
Product CodeNFB
Date Received2015-09-18
Returned To Mfg2015-08-28
Model NumberOM-501
OperatorPATIENT FAMILY MEMBER OR FRIEND
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerINOVO, INC
Manufacturer Address401 LEONARD BLVD. N. LEHIGH ACRES FL 33971 US 33971


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2015-09-18

© 2026 FDA.report