[26613691] Device not returned to manufacturer.
Patient Sequence No: 1, Text Type: N, H10


[26613692] A customer called in complaining of a contaminate on (b)(4),chromid(tm) chromid cps plates, lot 1004145580 discovered after incubation. Bacterium contamination is presenting as a brownish-yellow unidentified colony. Customer reports no patient results were affected. An internal investigation has been launched using retained plates.
Patient Sequence No: 1, Text Type: D, B5


[37853606] Review of qc batch records indicate conformance to specifications. During the investigation, sixty (60) chromid tm cps elite agar plates were tested; the plates were associated with three (3) different timeframes of the manufacturing process for the identified batch (11:25, 16:24 and 19:40). The plates were not inoculated. The plates were incubated for five (5) days at 33-37? C. Results conformed to specifications for contamination; zero (0) contaminations observed in 60 plates. There is no evidence of a contamination issue with the chromid tm cps elite agar product.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3002769706-2015-00096
MDR Report Key5091648
Date Received2015-09-21
Date of Report2015-08-27
Date Mfgr Received2015-08-27
Device Manufacturer Date2015-07-09
Date Added to Maude2015-09-21
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. RYAN LEMELLE
Manufacturer Street595 ANGLUM ROAD
Manufacturer CityHAZELWOOD MO 63042
Manufacturer CountryUS
Manufacturer Postal63042
Manufacturer Phone3147318582
Manufacturer G1BIOMERIEUX SA
Manufacturer StreetCHEMIN DE L ORME
Manufacturer CityMARCY L ETOILE, RHONE 69280
Manufacturer CountryFR
Manufacturer Postal Code69280
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCHROMID CPS ELITE TRANSLUCENT 100 PLT
Generic NameCHROMID CPS ELITE TRANSLUCENT 100 PLT
Product CodeJXA
Date Received2015-09-21
Catalog Number416172
Lot Number1004145580
Device Expiration Date2015-11-04
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerBIOMERIEUX SA
Manufacturer AddressCHEMIN DE L ORME MARCY L ETOILE, RHONE 69280 FR 69280


Patients

Patient NumberTreatmentOutcomeDate
10 2015-09-21

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.