VEST AIR PULSE GENERATOR P300413000

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2015-09-22 for VEST AIR PULSE GENERATOR P300413000 manufactured by Hill-rom Singapore.

Event Text Entries

[26683666] Hill-rom sent a replacement power cord. It is necessary for the vest? Airway clearance system to have an effective maintenance program. We recommend that you do annual preventive maintenance. The power cord is with the unit, and it is in good condition: the plug is a one-piece molded plug assembly, the assembly shows no signs of cracking, the plug molding around the blade is not discolored, and the blade is tight in the molding. A search of the hill-rom maintenance records did not show hill-rom performed any preventative maintenance on this unit. It is unknown if the account performs preventative maintenance on this unit. The account will replace the power cord to resolve the issue. Based on this information, no further action is required.
Patient Sequence No: 1, Text Type: N, H10


[26683667] Hill-rom received a report from the account stating the power cord sparked and left a burn mark on the end of the cord. The unit was located in the patients home. There was no patient/user injury reported. This report was filed in our complaint handling system as complaint (b)(4).
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3008145987-2015-00002
MDR Report Key5093721
Date Received2015-09-22
Date of Report2015-08-31
Date of Event2015-08-31
Date Mfgr Received2015-08-31
Date Added to Maude2015-09-22
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationPATIENT
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactTONY WERNER
Manufacturer Street1069 STATE ROUTE 46 EAST
Manufacturer CityBATESVILLE IN 47006
Manufacturer CountryUS
Manufacturer Postal47006
Manufacturer Phone8129312359
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameVEST AIR PULSE GENERATOR
Generic NameTHE VEST
Product CodeBYI
Date Received2015-09-22
Model NumberP300413000
OperatorLAY USER/PATIENT
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerHILL-ROM SINGAPORE
Manufacturer Address1 YISHUN AVENUE 7 SINGAPORE, NORTH EAST 768923 SN 768923


Patients

Patient NumberTreatmentOutcomeDate
10 2015-09-22