H. PYLORI STAT - PAK HP133

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2004-12-07 for H. PYLORI STAT - PAK HP133 manufactured by Chembio Diagnostic Systems, Inc..

Event Text Entries

[26967360] The product lot in question was bulk material for further packaging by (b)(4). Eval of device showed that it met release criteria.
Patient Sequence No: 1, Text Type: N, H10


[26967361] On (b)(6) 2004, a distributor/repackager. (b)(4), emailed saying that hp111703/1 of h. Pylori stat- pak has given 18 negatives in a suspect population. Three of these were confirmed positive by elisa. There was no involvement in any death or serious injury.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2431980-2004-00001
MDR Report Key5101371
Date Received2004-12-07
Date of Report2004-12-06
Date of Event2004-10-13
Date Mfgr Received2004-10-13
Device Manufacturer Date2003-11-17
Date Added to Maude2015-09-24
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationRISK MANAGER
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer Street3661 HORSEBLOCK RD.
Manufacturer CityMEDFORD NY 11763
Manufacturer CountryUS
Manufacturer Postal11763
Manufacturer Phone6319241135
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameH. PYLORI STAT - PAK
Generic NameIVD. RAPID DETECTION OF AB TO H. PYLORI
Product CodeOWF
Date Received2004-12-07
Returned To Mfg2004-11-22
Catalog NumberHP133
Lot NumberHP111703/1
Device Expiration Date2005-10-31
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerCHEMBIO DIAGNOSTIC SYSTEMS, INC.
Manufacturer AddressMEDFORD NY US


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2004-12-07