WATERPIK WATER FLOSSER WP-360

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a consumer report with the FDA on 2015-09-24 for WATERPIK WATER FLOSSER WP-360 manufactured by Water Pik, Inc..

Event Text Entries

[26683306] Unit in question has not yet been returned for testing and evaluation.
Patient Sequence No: 1, Text Type: N, H10


[26683307] Per the customer.... Father claims the (b)(6) daughter injured the roof of her mouth while flossing. Says daughter has used her wp-360 for some time with the high pressure tip and has had no problems. Suddenly the pressure became much higher than usual and caused small abrasions on the roof of her mouth. Daughter has braces, did not receive medical attention and abrasions healed quickly. Customer requested a replacement unit.
Patient Sequence No: 1, Text Type: D, B5


[31706083] Returned unit tested and found to be within the normal operating pressure range and was operating as designed. There is no known situation where pressure can temporarily increase. Most likely situation is that customer aimed water jet at soft tissue instead of gum area and caused pain. This is warned against in the product instructions for use.
Patient Sequence No: 1, Text Type: N, H10


[31706084] Per the customer.... Father claims the (b)(6) year old daughter injured the roof of her mouth while flossing. Says daughter has used her wp-360 for some time with the high pressure tip and has had no problems. Suddenly the pressure became much higher than usual and caused small abrasions on the roof of her mouth. Daughter has braces, did not receive medical attention and abrasions healed quickly. Customer requested a replacement unit.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number0001712259-2015-00006
MDR Report Key5101391
Report SourceCONSUMER
Date Received2015-09-24
Date of Report2015-11-19
Date of Event2015-09-23
Date Mfgr Received2015-09-23
Device Manufacturer Date2012-03-16
Date Added to Maude2015-09-24
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. ROBERT VANDER VLIET
Manufacturer Street1730 EAST PROSPECT ROAD
Manufacturer CityFORT COLLINS CO 805530001
Manufacturer CountryUS
Manufacturer Postal805530001
Manufacturer Phone9702217063
Manufacturer G1WATER PIK, INC.
Manufacturer Street1730 EAST PROSPECT ROAD
Manufacturer CityFORT COLLINS CO 805530001
Manufacturer CountryUS
Manufacturer Postal Code805530001
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameWATERPIK WATER FLOSSER
Generic NameORAL IRRIGATOR
Product CodeEFS
Date Received2015-09-24
Returned To Mfg2015-10-15
Model NumberWP-360
Lot NumberKL120316GA
OperatorLAY USER/PATIENT
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerWATER PIK, INC.
Manufacturer Address1730 EAST PROSPECT ROAD FORT COLLINS CO 805530001 US 805530001


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2015-09-24