MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2015-09-24 for TRIGLYCERIDES 05171407190 manufactured by Roche Diagnostics.
[26729566]
The triglyceride reagent lot number used was 567107 or 610229. The expiration dates were requested, but were not provided. This event occurred in (b)(6).
Patient Sequence No: 1, Text Type: N, H10
[26729567]
The customer received questionable low triglyceride results for two samples from one patient that were not accompanied by a data flag. The samples were tested on cobas 8000 c702 analyzer serial number (b)(4). For sample "(b)(4)" a result of 1. 39 mmol/l was generated with no flag. This result was reported outside the laboratory. A therapeutic insulin infusion was discontinued prematurely due to the result. This resulted in a delay in the patient's recovery. The patient's current condition was "recovered". No adverse event was reported. The sample was repeated multiple times and the repeat result of 48. 6 mmol/l was believed to be correct. For sample "(b)(4)" a result of 1. 42 mmol/l was generated with no flag. The sample was repeated multiple times. Results of 46. 29 mmol/l and 44. 76 mmol/l were generated. Refer to the attachment to the medwatch for all results for the patient.
Patient Sequence No: 1, Text Type: D, B5
Report Number | 1823260-2015-04215 |
MDR Report Key | 5101447 |
Report Source | FOREIGN,HEALTH PROFESSIONAL,U |
Date Received | 2015-09-24 |
Date of Report | 2015-09-24 |
Date of Event | 2015-06-18 |
Date Mfgr Received | 2015-09-04 |
Date Added to Maude | 2015-09-24 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | NA MICHAEL LESLIE |
Manufacturer Street | 9115 HAGUE ROAD NA |
Manufacturer City | INDIANAPOLIS IN 46250 |
Manufacturer Country | US |
Manufacturer Postal | 46250 |
Manufacturer Phone | 3175214343 |
Manufacturer G1 | ROCHE DIAGNOSTICS GMBH |
Manufacturer Street | SANDHOFERSTRASSE 116 NA |
Manufacturer City | MANNHEIM (BADEN-WURTTEMBERG) 68305 |
Manufacturer Country | GM |
Manufacturer Postal Code | 68305 |
Single Use | 3 |
Previous Use Code | 3 |
Removal Correction Number | NA |
Event Type | 3 |
Type of Report | 3 |
Brand Name | TRIGLYCERIDES |
Generic Name | TRIGLYCERIDES GPO WITHOUT FREE GLYCEROL |
Product Code | CDT |
Date Received | 2015-09-24 |
Model Number | NA |
Catalog Number | 05171407190 |
Lot Number | NI |
ID Number | NA |
Operator | HEALTH PROFESSIONAL |
Device Availability | Y |
Device Eval'ed by Mfgr | R |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | ROCHE DIAGNOSTICS |
Manufacturer Address | 9115 HAGUE ROAD NA INDIANAPOLIS IN 46250 US 46250 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2015-09-24 |