TRIGLYCERIDES 05171407190

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2015-09-24 for TRIGLYCERIDES 05171407190 manufactured by Roche Diagnostics.

Event Text Entries

[26729566] The triglyceride reagent lot number used was 567107 or 610229. The expiration dates were requested, but were not provided. This event occurred in (b)(6).
Patient Sequence No: 1, Text Type: N, H10


[26729567] The customer received questionable low triglyceride results for two samples from one patient that were not accompanied by a data flag. The samples were tested on cobas 8000 c702 analyzer serial number (b)(4). For sample "(b)(4)" a result of 1. 39 mmol/l was generated with no flag. This result was reported outside the laboratory. A therapeutic insulin infusion was discontinued prematurely due to the result. This resulted in a delay in the patient's recovery. The patient's current condition was "recovered". No adverse event was reported. The sample was repeated multiple times and the repeat result of 48. 6 mmol/l was believed to be correct. For sample "(b)(4)" a result of 1. 42 mmol/l was generated with no flag. The sample was repeated multiple times. Results of 46. 29 mmol/l and 44. 76 mmol/l were generated. Refer to the attachment to the medwatch for all results for the patient.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1823260-2015-04215
MDR Report Key5101447
Report SourceFOREIGN,HEALTH PROFESSIONAL,U
Date Received2015-09-24
Date of Report2015-09-24
Date of Event2015-06-18
Date Mfgr Received2015-09-04
Date Added to Maude2015-09-24
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactNA MICHAEL LESLIE
Manufacturer Street9115 HAGUE ROAD NA
Manufacturer CityINDIANAPOLIS IN 46250
Manufacturer CountryUS
Manufacturer Postal46250
Manufacturer Phone3175214343
Manufacturer G1ROCHE DIAGNOSTICS GMBH
Manufacturer StreetSANDHOFERSTRASSE 116 NA
Manufacturer CityMANNHEIM (BADEN-WURTTEMBERG) 68305
Manufacturer CountryGM
Manufacturer Postal Code68305
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameTRIGLYCERIDES
Generic NameTRIGLYCERIDES GPO WITHOUT FREE GLYCEROL
Product CodeCDT
Date Received2015-09-24
Model NumberNA
Catalog Number05171407190
Lot NumberNI
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerROCHE DIAGNOSTICS
Manufacturer Address9115 HAGUE ROAD NA INDIANAPOLIS IN 46250 US 46250


Patients

Patient NumberTreatmentOutcomeDate
10 2015-09-24

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.