NAVIOS FLOW CYTOMETER SYSTEM A52103

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,user facility report with the FDA on 2015-09-24 for NAVIOS FLOW CYTOMETER SYSTEM A52103 manufactured by Beckman Coulter Ireland.

Event Text Entries

[26727564] A field service engineer (fse) evaluated the gallios instrument at the customer's site. The fse found that an element on the tec slave control board was burnt (charred). Images submitted confirm the event. The board will be replaced when the replacement board is received. (b)(4).
Patient Sequence No: 1, Text Type: N, H10


[26727685] The customer reported a burning smell coming from a gallios flow cytometer system while restarting the instrument. There was no smoke, sparks, arcing or flames reported by the customer. The fire department was not called and there was no death, injury nor need for medical attention as a result of this event. Erroneous patient results were not generated and there was no change or effect to patient treatment in connection to the event. There was no impact to patient results and controls. This event occurred on the gallios device, which is marketed for research use only (ruo). This report is being filed for similar device, the navios flow cytometer system.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1061932-2015-01509
MDR Report Key5102210
Report SourceFOREIGN,USER FACILITY
Date Received2015-09-24
Date of Report2015-09-07
Date of Event2015-09-07
Date Mfgr Received2015-09-07
Device Manufacturer Date2013-12-01
Date Added to Maude2015-09-24
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. MIRANDA HOLLAND
Manufacturer Street11800 SW 147TH AVENUE, 32-S08
Manufacturer CityMIAMI FL 331962031
Manufacturer CountryUS
Manufacturer Postal331962031
Manufacturer Phone3053802031
Manufacturer G1BECKMAN COULTER
Manufacturer Street11800 SW 147TH AVENUE
Manufacturer CityMIAMI FL 33196
Manufacturer CountryUS
Manufacturer Postal Code33196
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameNAVIOS FLOW CYTOMETER SYSTEM
Generic NameFLOW CYTOMETRIC REAGENTS AND ACCESSORIES
Product CodeOYE
Date Received2015-09-24
Model NumberNA
Catalog NumberA52103
Lot NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerBECKMAN COULTER IRELAND
Manufacturer AddressLISMEEHAN O'CALLAGHAN'S MILLS, COUNTY CLARE EI


Patients

Patient NumberTreatmentOutcomeDate
10 2015-09-24