MILLENIUM RO SYSTEM

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional,user faci report with the FDA on 2015-09-25 for MILLENIUM RO SYSTEM manufactured by Mar Cor Purification.

Event Text Entries

[26758087] On (b)(6), dialysis clinic accidentally connected the product water line to the disinfection port of the portable ro system (ports are color coded, unit manufactured prior to 2005). One patient was dialyzed with de-chlorinated water that had not been treated by the reverse osmosis water system. Clinic tested the water and found elevated sodium and fluoride levels. No patient or worker injuries were reported. This complaint will continue to be maintained within the mar cor complaint system.
Patient Sequence No: 1, Text Type: N, H10


[26758088] Patient was dialyzed utilizing de-chlorinated water that has not been processed by the portable ro system. Biomed manager was not aware of any patient issues (one patient).
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3019131-2015-00004
MDR Report Key5104050
Report SourceHEALTH PROFESSIONAL,USER FACI
Date Received2015-09-25
Date of Report2015-09-25
Date of Event2015-08-27
Date Mfgr Received2015-08-27
Device Manufacturer Date2004-12-31
Date Added to Maude2015-09-25
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. MEGAN DICKEY
Manufacturer Street14550 28TH AVENUE NORTH
Manufacturer CityMINNEAPOLIS MN 55447
Manufacturer CountryUS
Manufacturer Postal55447
Manufacturer Phone7635533300
Manufacturer G1MAR COR PURIFICATION
Manufacturer Street14550 28TH AVENUE NORTH
Manufacturer CityMINNEAPOLIS MN 55447
Manufacturer CountryUS
Manufacturer Postal Code55447
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameMILLENIUM RO SYSTEM
Generic NameWATER PURIFICATION SYSTEMS AND PORTABLE REVERSE OSMOSIS SYSTEMS
Product CodeFIP
Date Received2015-09-25
Model NumberMILLENIUM
Device AvailabilityY
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerMAR COR PURIFICATION
Manufacturer Address14550 28TH AVENUE NORTH MINNEAPOLIS MN 55447 US 55447


Patients

Patient NumberTreatmentOutcomeDate
10 2015-09-25

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.