MEDICAL GAS FLOW METER F56-210D2 *

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2003-12-01 for MEDICAL GAS FLOW METER F56-210D2 * manufactured by Flotec, Inc..

Event Text Entries

[310569] After disconnecting the oxygen flow meter to record the flow rate, the respiratory therapist was attempting to reconnect the meter to the oxygen wall outlet, a jet of water sprayed the therapist in the face. The water came from the humidifier because the flowmeter delivered a sudden burst of gas when the flowmeter was plugged into the wall. The humidifier was connected to the flow meter prior to the pulgging into the wall. The flowmeter was the problem. They cannot find a consistent condition that caused the flowmeter to do this. If the flowmeter is set to zero flow when plugged -in then adjusted to deliver oxygen flowrate, the flowmeter doesn't seem to fail. The wall outlet was find. No water reached the pt. Device usage problem: device malfunction - that is, the device did not do what it was supposed to do.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number510440
MDR Report Key510440
Date Received2003-12-01
Date of Report2003-12-01
Date of Event2003-12-01
Date Added to Maude2004-02-10
Event Key0
Report Source CodeUser Facility report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Single Use0
Previous Use Code0
Event Type3
Type of Report3

Device Details

Brand NameMEDICAL GAS FLOW METER
Generic NameGAS FLOW METER
Product CodeCCN
Date Received2003-12-01
Model NumberF56-210D2
Catalog Number*
Lot Number*
ID Number*
Device AvailabilityY
Device Age6 MO
Implant FlagN
Date Removed*
Device Sequence No1
Device Event Key499390
ManufacturerFLOTEC, INC.
Manufacturer Address7625 W NEW YORK ST INDIANAPOLIS IN 462144911 US


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2003-12-01