DISPOSABLE SCALP NEEDLE ELECTRODES 9013R0313

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a distributor report with the FDA on 2015-09-29 for DISPOSABLE SCALP NEEDLE ELECTRODES 9013R0313 manufactured by Alpine Biomed Aps.

Event Text Entries

[26973095] Registered internally as a (b)(4)) for investigative purposes. The contract manufacturer sp medical have also been issued a supplier corrective action report (scar) to provide details and results of their investigation into separation of the needles and all resulting corrective and preventative action ((b)(4)) the device history record for this batch was provided by (b)(6) medical and reviewed. In process testing and inspections were reviewed. All samples taken for tests gave acceptable result. No issues were noted. No rework or sort performed on this lot. No ncr's noted for this lot.
Patient Sequence No: 1, Text Type: N, H10


[26973096] Customer reported finding a broken needle on opening a box of scalp needles and another needle from the same box broke while in the patients head. The needle came apart from the hub during use, leaving the needle still inserted in the patients head.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3004827015-2015-00001
MDR Report Key5109965
Report SourceDISTRIBUTOR
Date Received2015-09-29
Date of Report2015-09-28
Date of Event2015-09-11
Date Mfgr Received2015-09-12
Device Manufacturer Date2015-02-24
Date Added to Maude2015-09-29
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR MICHAEL GALVIN
Manufacturer StreetIDA BUSINESS PARK GORT
Manufacturer CityGALWAY,
Manufacturer CountryEI
Manufacturer Phone91647451
Manufacturer G1SP MEDICAL A/S
Manufacturer StreetM 4653 KARISE
Manufacturer City4653
Manufacturer CountryDA
Manufacturer Postal Code4653
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameDISPOSABLE SCALP NEEDLE ELECTRODES
Generic NameELECTRODE
Product CodeGXZ
Date Received2015-09-29
Model Number9013R0313
Catalog Number9013R0313
Lot Number60550
Device Expiration Date2019-03-31
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerALPINE BIOMED APS
Manufacturer AddressTONSBAKKEN 16-18 DK - 2740 SKOVLUNDE, DENMARK DK-2740 DA DK-2740


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2015-09-29