55 AMP HIGH EFFICIENCY AC COMPUTER CART 1781707

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2015-09-30 for 55 AMP HIGH EFFICIENCY AC COMPUTER CART 1781707 manufactured by Rubbermaid Medical Solutions Inc./healthcare Products Inc..

Event Text Entries

[27166397]
Patient Sequence No: 1, Text Type: N, H10


[27166398] There are approximately 170 rubbermaid m40 mobile carts that are used in all areas of our institution. During the last 8 days, we have experience 3 fires associated with these devices. All fires were experienced in the critical care units. Upon opening the units after the fire, it appears that some of the components on the circuit board are getting really hot and catching on fire. At this time, we are not sure if the actual fire is a result of a faulty electrical component such as a capacitor/transistor or if it is the result of heat build up due to excessive lint in the power supply area. All failed units were manufactured in 2011.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number5112576
MDR Report Key5112576
Date Received2015-09-30
Date of Report2015-09-29
Date of Event2015-09-21
Report Date2015-09-29
Date Reported to FDA2015-09-29
Date Reported to Mfgr2015-09-29
Date Added to Maude2015-09-30
Event Key0
Report Source CodeUser Facility report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag0
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand Name55 AMP HIGH EFFICIENCY AC COMPUTER CART
Generic NameMEDICAL DEVICE DATA SYSTEM
Product CodeLNX
Date Received2015-09-30
Model Number1781707
Device Availability*
Device Age3 YR
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerRUBBERMAID MEDICAL SOLUTIONS INC./HEALTHCARE PRODUCTS INC.
Manufacturer Address8935 NORTHPOINT EXECUTIVE DRIVE HUNTERSVILLE NC 28078 US 28078


Patients

Patient NumberTreatmentOutcomeDate
10 2015-09-30