ANCHOR BOLT LSBK2-_X-04

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional report with the FDA on 2015-10-01 for ANCHOR BOLT LSBK2-_X-04 manufactured by Ad-tech Medical Instrument Corp..

Event Text Entries

[27247609] As stated in the event description section, this issue only occurs during the lateral approach to the hippocampus and amygdala. The bone that is drilled into prior to performing the lateral approach is called the squamous temporal bone. This bone is much thinner and can be only 2-3 mm thick. The complaint investigation team is currently in the process of investigating the potential malfunction.
Patient Sequence No: 1, Text Type: N, H10


[27247610] On (b)(6) 2015, customer feedback was received related to one of ad-tech's anchor bolt. Specifically, the customer stated that during approximately 50% of the surgeries involving the lateral approach to the hippocampus and amygdala, they have experienced that the anchor bolt does not hold well when the entry point is over a "thin" squamous temporal bone. The anchor bolt gets very little purchase and is very loose, potentially causing the electrode to move. The customer also noted that if the squamous temporal bone is thick, the anchor bolt holds well. This is considered a potential malfunction; however, there have been no reports of patient harm when this type of instance occurs.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2183456-2015-00008
MDR Report Key5117630
Report SourceHEALTH PROFESSIONAL
Date Received2015-10-01
Date of Report2015-09-01
Date Mfgr Received2015-09-01
Date Added to Maude2015-10-01
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS KATHLEEN BARLOW
Manufacturer Street1901 WILLIAM STREET
Manufacturer CityRACINE WI 53404
Manufacturer CountryUS
Manufacturer Postal53404
Manufacturer Phone2626341555
Manufacturer G1AD-TECH MEDICAL INSTRUMENT CORP.
Manufacturer Street1901 WILLIAM STREET
Manufacturer CityRACINE WI 53404
Manufacturer CountryUS
Manufacturer Postal Code53404
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameANCHOR BOLT
Generic NameANCHOR BOLT
Product CodeGCZ
Date Received2015-10-01
Catalog NumberLSBK2-_X-04
Lot NumberUNKNOWN
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerAD-TECH MEDICAL INSTRUMENT CORP.
Manufacturer Address1901 WILLIAM STREET RACINE WI 53404 US 53404


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2015-10-01

© 2026 FDA.report