ALERE CHOLESTECH LDX MULTIANALYTE CONTROLS 88773

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2015-10-02 for ALERE CHOLESTECH LDX MULTIANALYTE CONTROLS 88773 manufactured by Alere San Diego, Inc..

Event Text Entries

[27650076] Mdr is being filed due to field action: ref-2027969-08/25/15-001c. Investigation pending.
Patient Sequence No: 1, Text Type: N, H10


[27650077] Customer reported that controls were outside of the expected range - high. Results were as follows: tc high two times for l1. Tc=160, 157 (108-152); trg=132 (129-191). The results were within the corrected expected values. Technical service reviewed factors and informed customer of notification; provided corrected expected values ranges and emailed the notification to the customer.
Patient Sequence No: 1, Text Type: D, B5


[33032345] Control lot c3091a was associated with field action per internal (b)(4) for incorrect expected value card provided with this control lot with the incorrect concentration assignment ranges for the total cholesterol and triglycerides analytes. Investigation will be performed under internal (b)(4).
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2027969-2015-00825
MDR Report Key5125624
Date Received2015-10-02
Date of Report2015-09-04
Date of Event2015-09-04
Date Mfgr Received2015-10-08
Date Added to Maude2015-10-02
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationDIETICIAN
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactYA-LING KING
Manufacturer Street9975 SUMMERS RIDGE RD.
Manufacturer CitySAN DIEGO CA 92121
Manufacturer CountryUS
Manufacturer Postal92121
Manufacturer Phone8588052084
Manufacturer G1ALERE SAN DIEGO, INC.
Manufacturer Street9975 SUMMERS RIDGE RD.
Manufacturer CitySAN DIEGO CA 92121
Manufacturer CountryUS
Manufacturer Postal Code92121
Single Use3
Previous Use Code3
Removal Correction Number2027969-08/25/15-001C
Event Type3
Type of Report3

Device Details

Brand NameALERE CHOLESTECH LDX MULTIANALYTE CONTROLS
Generic NameMULTI-ANALYTE CONTROLS
Product CodeJJY
Date Received2015-10-02
Model Number88773
Lot NumberC3091A
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerALERE SAN DIEGO, INC.
Manufacturer Address9975 SUMMERS RIDGE RD. SAN DIEGO CA 92121 US 92121


Patients

Patient NumberTreatmentOutcomeDate
10 2015-10-02

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.