RESQCPR SYSTEM 12-0823-000

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional report with the FDA on 2015-10-08 for RESQCPR SYSTEM 12-0823-000 manufactured by .

Event Text Entries

[27809226] Preliminary findings indicate that the hole in the clam shell lower handle is undersized and out of specification. Therefore when the transfer spring subassembly is secured into the clamshell lower assembly, a potential exists for an interference fit to occur between these parts. Due to this interference fit, when the resqpump is exercised during cpr the transfer spring will wear or distort the hole in the lower clamshell handle, reducing the friction fit between the transfer spring subassembly and lower clamshell. The result of this wear or distortion is that the unit allows the transfer spring subassembly to translate up and down in the socket in the clamshell, coinciding with compressions and decompressions and may develop a "clicking" sound. This motion of the transfer spring subassembly may affect the calibration of the force gauge on the device as the two components progressively wear together.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3003477173-2015-00003
MDR Report Key5134648
Report SourceHEALTH PROFESSIONAL
Date Received2015-10-08
Date of Report2015-10-01
Date of Event2015-08-31
Date Added to Maude2015-10-08
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional0
Initial Report to FDA0
Report to FDA3
Event Location3
Manufacturer ContactDIANE HOWELL
Manufacturer Street1905 COUNTY ROAD C
Manufacturer CityROSEVILLE MN 55113
Manufacturer CountryUS
Manufacturer Postal55113
Manufacturer Phone6514035600
Manufacturer G1SCIENTIFIC MOLDING CORPORATION
Manufacturer Street330 SMC DRIVE
Manufacturer CitySOMERSET WI 54025
Manufacturer CountryUS
Manufacturer Postal Code54025
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameRESQCPR SYSTEM
Generic NameRESQPUMP
Product CodePIZ
Date Received2015-10-08
Returned To Mfg2015-09-18
Catalog Number12-0823-000
Lot Number44598758
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0


Patients

Patient NumberTreatmentOutcomeDate
10 2015-10-08

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.