VALSURE NEUTRAL DETERGENT

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a user facility report with the FDA on 2015-10-08 for VALSURE NEUTRAL DETERGENT manufactured by Steris Corporation.

Event Text Entries

[27830831] The user facility used valsure neutral detergent to soak their instruments prior to use in patient procedures. After the procedures were completed, two patients reported symptoms of toxic anterior segment syndrome (tass). The user contacted steris to inquire about valsure as a part of their investigation for the reported event as this was recently added to their process for cleaning. The user facility declined to provide additional information regarding the reported injury or current status of the patients subject of the reported event. A steris account manager spoke with the user facility regarding the use of valsure neutral detergent and identified the user facility was not following the instructions for use as indicated on the product's label. The cleaning solution was diluted to 1/8 oz per 1 gallon of water which is below the recommended dosing for the cleaning solution. The instructions for use for valsure neutral detergent state, "dilute at 1-3 ounces per gallon, depending on water quality and soil load. " after reviewing the instructions for use, the user facility acknowledged that they do not believe that valsure neutral detergent contributed to the reported injury. Steris has performed in-service training with user facility about the proper cleaning techniques and dosages for valsure neutral detergent to ensure proper cleaning of their surgical instruments.
Patient Sequence No: 1, Text Type: N, H10


[27830832] The user facility contacted steris to inquire about the proper cleaning procedure for cleaning instruments with valsure neutral detergent.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1937531-2015-00010
MDR Report Key5135572
Report SourceUSER FACILITY
Date Received2015-10-08
Date of Report2015-10-07
Date of Event2015-09-10
Date Mfgr Received2015-09-10
Date Added to Maude2015-10-08
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. KATHRYN CADORETTE
Manufacturer Street5960 HEISLEY ROAD
Manufacturer CityMENTOR OH 44060
Manufacturer CountryUS
Manufacturer Postal44060
Manufacturer Phone4403927231
Manufacturer G1STERIS CORPORATION
Manufacturer Street7501 PAGE AVENUE
Manufacturer CityST. LOUIS MO 63133
Manufacturer CountryUS
Manufacturer Postal Code63133
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameVALSURE NEUTRAL DETERGENT
Generic NameDETERGENT
Product CodeFLG
Date Received2015-10-08
Lot Number274714 14A5.
OperatorHEALTH PROFESSIONAL
Device Availability*
Device Eval'ed by Mfgr*
Device Sequence No1
Device Event Key0
ManufacturerSTERIS CORPORATION
Manufacturer Address7501 PAGE AVENUE ST. LOUIS MO 63133 US 63133


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2015-10-08

© 2026 FDA.report