BIPOLAR WORKING ELEMENT 8680.224

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,user f report with the FDA on 2015-10-08 for BIPOLAR WORKING ELEMENT 8680.224 manufactured by Richard Wolf Medical Instruments Corp (rwmic).

Event Text Entries

[27766899] Investigation revealed device was missing screw and thumb ring. Device also had an electrical short. In addition, the device is over two years old and had never been submitted to rwmic for any repairs or routine maintenance. Manufacture date: 12/2012. Purchase date: (b)(6) 2013. Additional information requested concerning the event, no response as of 10/07/2015. Rwmic considers this matter closed. However, in the event we receive additional information, we will provide fda with follow-up information.
Patient Sequence No: 1, Text Type: N, H10


[27766900] During a case the device became hot and doctor felt a shock. After case completed doctor noticed burn on finger.
Patient Sequence No: 1, Text Type: D, B5


[28869216] User facility submitted additional information on 10/21/2015. Added patient information and initial reporter information.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1418479-2015-00034
MDR Report Key5135923
Report SourceCOMPANY REPRESENTATIVE,USER F
Date Received2015-10-08
Date of Report2015-10-21
Date of Event2015-09-09
Date Mfgr Received2015-10-21
Device Manufacturer Date2012-12-01
Date Added to Maude2015-10-08
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationSERVICE PERSONNEL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS DAWN CLARK
Manufacturer Street353 CORPORTATE WOODS PARKWAY
Manufacturer CityVERNON HILLS IL 600613110
Manufacturer CountryUS
Manufacturer Postal600613110
Manufacturer Phone8003239653
Manufacturer G1RICHARD WOLF MEDICAL INSTRUMENTS CORP (RWMIC)
Manufacturer Street353 CORPORTATE WOODS PARKWAY
Manufacturer CityVERNON HILLS IL 600613110
Manufacturer CountryUS
Manufacturer Postal Code600613110
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameBIPOLAR WORKING ELEMENT
Generic NameWORKING ELEMENT
Product CodeFDC
Date Received2015-10-08
Returned To Mfg2015-09-23
Model Number8680.224
Catalog Number8680.224
Lot Number179D12
OperatorPHYSICIAN
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerRICHARD WOLF MEDICAL INSTRUMENTS CORP (RWMIC)
Manufacturer Address353 CORPORTATE WOODS PARKWAY VERNON HILLS IL 600613110 US 600613110


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2015-10-08

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.