AMS ACTICON NEOSPHINCTER

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional report with the FDA on 2015-10-08 for AMS ACTICON NEOSPHINCTER manufactured by American Medical Systems (mn).

Event Text Entries

[27768547] Additional information: cuff: catalog #: 72401982; expiration date: 4/22/2019; serial #: (b)(4); manufacture date: 4/2014. Pump: catalog #: 72402287; expiration date: 11/20/2015; serial #: (b)(4), manufacture date: 11/2014. Balloon: catalog #: 72402106; expiration date: 4/17/2018; serial #: (b)(4), manufacture date: 5/2013. The explanted action bowel sphincter was visually inspected and functionally tested. The pump performed within specifications. The balloon pressure tested at 99. 7 cmh2o. It was originally rated at 101-110 cmh2o. There was a leak in the cuff that was the result of a sharp instrument.
Patient Sequence No: 1, Text Type: N, H10


[27768548] It was reported the patient had his action bowel sphincter removed due to erosion and infection. No additional patient complications were reported in relation to this event.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2183959-2015-00422
MDR Report Key5136596
Report SourceHEALTH PROFESSIONAL
Date Received2015-10-08
Date of Report2015-09-14
Date of Event2015-08-31
Date Mfgr Received2015-09-14
Date Added to Maude2015-10-08
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMRS. SHARON ZURN
Manufacturer Street10700 BREN ROAD W
Manufacturer CityMINNETONKA MN 55343
Manufacturer CountryUS
Manufacturer Postal55343
Manufacturer Phone9529306000
Manufacturer G1AMERICAN MEDICAL SYSTEMS (MN)
Manufacturer Street10700 BREN ROAD W
Manufacturer CityMINNETONKA MN 55343
Manufacturer CountryUS
Manufacturer Postal Code55343
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameAMS ACTICON NEOSPHINCTER
Generic NameIMPLANTED FECAL INCONTINENCE DEVICE
Product CodeMIP
Date Received2015-10-08
Returned To Mfg2015-09-24
OperatorLAY USER/PATIENT
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerAMERICAN MEDICAL SYSTEMS (MN)
Manufacturer Address10700 BREN ROAD W MINNETONKA MN 55343 US 55343


Patients

Patient NumberTreatmentOutcomeDate
101. Hospitalization; 2. Required No Informationntervention 2015-10-08