MAYFIELD 4-0-A-2020 A2020

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a other report with the FDA on 2004-07-22 for MAYFIELD 4-0-A-2020 A2020 manufactured by Integra Lifesciences Corporation.

Event Text Entries

[27941852] Integra lifesciences corporation detected a fracturing condition during the inspection process of the skull pin. Radiolucent skull pins suspected of having this condition was traced to lot# 42358.
Patient Sequence No: 1, Text Type: N, H10


[27941853] Lot #42358 is suspected of having skull pins that possibly could fracture when under compression. There have been no field reported injuries or incidents involving the radiolucent skull pins. Integra lifesciences is voluntarily recalling all 40a2020 skull pins with lot #42358.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3004608878-2004-00007
MDR Report Key5137025
Report SourceOTHER
Date Received2004-07-22
Date Mfgr Received2004-06-21
Device Manufacturer Date2004-06-04
Date Added to Maude2015-10-08
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactGENE WEBBER
Manufacturer Street4900 CHARLEMAR DR.,
Manufacturer CityCINCINNATI OH 45227
Manufacturer CountryUS
Manufacturer Postal45227
Manufacturer Phone5132729113
Single Use3
Previous Use Code3
Removal Correction Number2004-00007R
Event Type3
Type of Report0

Device Details

Brand NameMAYFIELD
Generic NameSKULL PINS
Product CodeHAX
Date Received2004-07-22
Model Number4-0-A-2020
Catalog NumberA2020
Lot Number42358
Device Availability*
Device Age2 MO
Device Eval'ed by Mfgr*
Device Sequence No1
Device Event Key0
ManufacturerINTEGRA LIFESCIENCES CORPORATION
Manufacturer Address4900 CHARLEMAR DR., CINCINNATI OH 45227 US 45227


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2004-07-22

© 2026 FDA.report