EASYSTREAM (TM) 414352

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2015-10-09 for EASYSTREAM (TM) 414352 manufactured by Biomerieux Sa.

Event Text Entries

[28571009] Summary: customer reported the easystream (tm), sn (b)(4) is not providing an indication when eluates have not been distributed correctly (not alarming). Customer reports that there has been a delay of up to 48 hours no specific details have been provided. An investigation has been initiated.
Patient Sequence No: 1, Text Type: D, B5


[33469428] Biom? Rieux internal investigation was conducted. The customer's issue was reproduced six (6) times in >3000 pipette attempts. Potential root causes, primary among them being excessive friction within the pipettor head, have been identified and analysis is being performed under corrective and preventive action (capa) record (b)(4). In the meantime, the following actions have occurred: easy stream instruments and pipettor heads in inventory at the international distribution center (idc) were returned to the supplier on (b)(4) 2015 for re-work, and subsequently received back for use on (b)(4) 2015. Product stop shipment (b)(4) was issued to subsidiaries and distributors 09 oct 2015. Biom? Rieux and the supplier defined a new quality control test for pipettor head performance verification. Field corrective action (b)(4) was issued to the field on 30 oct 2015 instructing replacement of the pipettor head at each customer site, using newly established quality control tests for performance confirmation.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3002769706-2015-00113
MDR Report Key5141314
Date Received2015-10-09
Date of Report2015-07-29
Date Mfgr Received2015-07-29
Date Added to Maude2015-10-09
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. RYAN LEMELLE
Manufacturer Street595 ANGLUM ROAD
Manufacturer CityHAZELWOOD MO 63042
Manufacturer CountryUS
Manufacturer Postal63042
Manufacturer Phone3147318582
Manufacturer G1BIOMERIEUX SA
Manufacturer StreetCHEMIN DE L ORME
Manufacturer CityMARCY L ETOILE, RHONE 69280
Manufacturer CountryFR
Manufacturer Postal Code69280
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameEASYSTREAM (TM)
Generic NameEASYSTREAM (TM
Product CodeJQW
Date Received2015-10-09
Catalog Number414352
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerBIOMERIEUX SA
Manufacturer AddressCHEMIN DE L ORME MARCY L ETOILE, RHONE 69280 FR 69280


Patients

Patient NumberTreatmentOutcomeDate
10 2015-10-09

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.