VITROS CHEMISTY PRODUCTS AMON SLIDES 1726926

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2015-10-15 for VITROS CHEMISTY PRODUCTS AMON SLIDES 1726926 manufactured by Ortho-clinical Diagnostics.

Event Text Entries

[28735503] The customer observed higher than expected vitros amon quality control results from vitros liquid performance verifier control fluid using vitros amon micro slides on a vitros 5,1 fs chemistry system. A user error related to the use of expired reagents is the likely assignable cause. In addition, an unidentified instrument issue cannot be ruled out as a contributing factor as the customer would not perform recommended testing to verify instrument performance.
Patient Sequence No: 1, Text Type: N, H10


[28735504] The customer observed higher than expected vitros amon quality control results from vitros liquid performance verifier control fluid using vitros amon micro slides on a vitros 5,1 fs chemistry system. Level 2 results: 247. 0 and 235. 0 umol/l vs. Expected 194. 2 umol/l. Biased results of the direction and magnitude observed may lead to inappropriate physician action. Although there were no reports of affected patient sample results, it cannot be confirmed that patient sample results were not affected and would not be affected if the event were to recur undetected. There were no allegations of patient harm. This report is number one of two mdr's for this event. Two 3500a forms are being submitted for this event as two devices were involved. (b)(4).
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1319809-2015-00109
MDR Report Key5153652
Date Received2015-10-15
Date of Report2015-10-15
Date of Event2015-09-17
Date Mfgr Received2015-09-18
Device Manufacturer Date2013-09-20
Date Added to Maude2015-10-15
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. JAMES A STEVENS
Manufacturer Street100 INDIGO CREEK DRIVE
Manufacturer CityROCHESTER NY 14626
Manufacturer CountryUS
Manufacturer Postal14626
Manufacturer Phone5854533000
Manufacturer G1ORTHO-CLINICAL DIAGNOSTICS
Manufacturer Street513 TECHNOLOGY BLVD.
Manufacturer CityROCHESTER NY 14652
Manufacturer CountryUS
Manufacturer Postal Code14652
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameVITROS CHEMISTY PRODUCTS AMON SLIDES
Generic NameIN-VITRO DIAGNOSTIC
Product CodeJID
Date Received2015-10-15
Catalog Number1726926
Lot Number1015-0233-7466
Device Expiration Date2015-03-01
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerORTHO-CLINICAL DIAGNOSTICS
Manufacturer Address100 INDIGO CREEK DRIVE ROCHESTER NY 14626 US 14626


Patients

Patient NumberTreatmentOutcomeDate
10 2015-10-15

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.