ROCHE ELECSYS ANTI-CCP IMMUNOASSAY 05031656190

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2015-10-20 for ROCHE ELECSYS ANTI-CCP IMMUNOASSAY 05031656190 manufactured by Roche Diagnostics.

Event Text Entries

[28805983] This event occurred in (b)(6).
Patient Sequence No: 1, Text Type: N, H10


[28805984] The customer received questionable anti-ccp results for one patient from an unknown roche analyzer. On (b)(6) 2015, the result was 478 u/ml, but the result by stacia, meblux test whose principle was cleia, was negative. Additionally, the patient's rf was negative. It was unclear if this testing was performed with the same sample. On (b)(6) 2015, the result was 453 u/ml, but the result for the same sample with by stacia, meblux test was negative. The rf result was negative and the anti-nuclear antibody (ana) was <40 folds. Information concerning if any erroneous result was reported outside the laboratory was requested, but was not provided. The patient was not adversely affected.
Patient Sequence No: 1, Text Type: D, B5


[33614111] A specific root cause could not be identified as the sample was not available for further investigation. A possible cause for the issue could have been an interference in the sample.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1823260-2015-04357
MDR Report Key5162018
Report SourceFOREIGN,HEALTH PROFESSIONAL,U
Date Received2015-10-20
Date of Report2015-12-16
Date of Event2015-06-08
Date Mfgr Received2015-10-02
Date Added to Maude2015-10-20
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactNA MICHAEL LESLIE
Manufacturer Street9115 HAGUE ROAD NA
Manufacturer CityINDIANAPOLIS IN 46250
Manufacturer CountryUS
Manufacturer Postal46250
Manufacturer Phone3175214343
Manufacturer G1ROCHE DIAGNOSTICS GMBH
Manufacturer StreetSANDHOFERSTRASSE 116 NA
Manufacturer CityMANNHEIM (BADEN-WURTTEMBERG) 68305
Manufacturer CountryGM
Manufacturer Postal Code68305
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameROCHE ELECSYS ANTI-CCP IMMUNOASSAY
Generic NameANTIBODIES, ANTI-CYCLIC CITRULLINATED PEPTIDE (CCP)
Product CodeNHX
Date Received2015-10-20
Model NumberNA
Catalog Number05031656190
Lot NumberASKU
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerROCHE DIAGNOSTICS
Manufacturer Address9115 HAGUE ROAD NA INDIANAPOLIS IN 46250 US 46250


Patients

Patient NumberTreatmentOutcomeDate
10 2015-10-20

© 2026 FDA.report