[29256684] (b)(6) 2015 09:12 am (gmt-4:00) added by (b)(6) ((b)(4)): a maquet field service technician evaluated the device and found that the root cause of this incident was a missing bolt. The bolt that holds the main pole to the wheels assembly could not be located after the event. However, scuff marks in the steel prove that the missing part was there initially. It has not been determined why the bolt was missing. The lucea led operating manual includes instructions to check annually that all retaining screws and all visible screws of the shaft and the arm are tightly fixed. Additionally, the users are warned that the light could fall over, and are requested to take care when moving the mobile stand. Component codes: (b)(4).
Patient Sequence No: 1, Text Type: N, H10


[29256705] The customer reported to maquet service that a part of a mobile surgical light disconnected during surgery when the or technician attempted to move it. When trying to prevent the light from falling to the floor, she was hit on the right shoulder. No information was provided regarding the outcome of the technician. No injury to patient was reported. Factory reference number: (b)(4).
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number9710055-2015-01020
MDR Report Key5166933
Date Received2015-10-21
Date of Report2015-09-23
Date of Event2015-09-23
Date Mfgr Received2015-09-23
Device Manufacturer Date2014-03-01
Date Added to Maude2015-10-21
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationBIOMEDICAL ENGINEER
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer StreetPARC DE LIM AVENUE DE LA POMME DE PI
Manufacturer CityORL 45074
Manufacturer CountryFR
Manufacturer Postal45074
Manufacturer G1MARIE-FRAN
Manufacturer StreetPARC DE LIM AVENUE DE LA POMME DE PI
Manufacturer CityORL 45074
Manufacturer CountryFR
Manufacturer Postal Code45074
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameLUCEA LED
Generic NameLIGHT, SURGICAL, FLOOR STANDING
Product CodeFSS
Date Received2015-10-21
Model NumberLCA50 MOBILE B
Catalog NumberLCA309005C
ID Number568604977
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeYR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerMAQUET SAS
Manufacturer AddressORL?ANS CEDEX 2 FR


Patients

Patient NumberTreatmentOutcomeDate
10 2015-10-21

© 2026 FDA.report