MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2015-10-21 for STRAIGHT MBLOC ACETABULAR SHELL INSERTER N/A 110003450 manufactured by Biomet Orthopedics.
[28903847]
Current information is insufficient to permit a conclusion as to the cause of the event. Review of device history records show that lot released with no recorded anomaly or deviation. Device availability - the device is reported to be available for evaluation; however, it has not been received by biomet orthopedics to date. In the event that the device is received and evaluated, a follow up report will be sent to the fda to provide results. There are warnings in the package insert that state that this type of event can occur: under care and handling of instruments, it states,? Surgical instruments and instrument cases are susceptible to damage for a variety of reasons including prolonged use, misuse, rough or improper handling. Care must be taken to avoid compromising their exacting performance.?
Patient Sequence No: 1, Text Type: N, H10
[28903848]
It was reported that patient underwent a total hip arthroplasty on (b)(6), 2015. During the procedure, the inserter would not disengage from the acetabular cup after impaction. As a result, the cup was removed and another inserter and cup were utilized to complete the procedure. A delay greater than 30 minutes occurred as a result.
Patient Sequence No: 1, Text Type: D, B5
[35318936]
This follow-up report is being filed to relay additional information, which was unknown at the time of the initial medwatch.
Patient Sequence No: 1, Text Type: N, H10
[36807529]
Examination of returned device found no evidence of product non-conformance. Review of the device confirmed the reported condition. Root cause of the event was most likely attributed to damage to the threads of either the inserter or the cup; however, a conclusive root cause could not be determined.
Patient Sequence No: 1, Text Type: N, H10
Report Number | 0001825034-2015-04399 |
MDR Report Key | 5166962 |
Date Received | 2015-10-21 |
Date of Report | 2016-01-20 |
Date of Event | 2015-09-30 |
Date Mfgr Received | 2016-01-20 |
Device Manufacturer Date | 2014-01-14 |
Date Added to Maude | 2015-10-21 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | MS. MEGAN HAAS |
Manufacturer Street | 56 E. BELL DRIVE |
Manufacturer City | WARSAW IN 46582 |
Manufacturer Country | US |
Manufacturer Postal | 46582 |
Manufacturer Phone | 5743726700 |
Manufacturer G1 | BIOMET ORTHOPEDICS |
Manufacturer Street | 56 E. BELL DRIVE |
Manufacturer City | WARSAW IN 46582 |
Manufacturer Country | US |
Manufacturer Postal Code | 46582 |
Single Use | 3 |
Previous Use Code | 3 |
Removal Correction Number | N/A |
Event Type | 3 |
Type of Report | 3 |
Brand Name | STRAIGHT MBLOC ACETABULAR SHELL INSERTER |
Generic Name | PROSTHESIS, HIP |
Product Code | LKD |
Date Received | 2015-10-21 |
Returned To Mfg | 2015-12-18 |
Model Number | N/A |
Catalog Number | 110003450 |
Lot Number | 856210 |
ID Number | N/A |
Operator | PHYSICIAN |
Device Availability | R |
Device Age | DA |
Device Eval'ed by Mfgr | R |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | BIOMET ORTHOPEDICS |
Manufacturer Address | 56 E. BELL DRIVE WARSAW IN 46582 US 46582 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 1. Other | 2015-10-21 |