CHROMID(TM) CPS ELITE 418284

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2015-10-24 for CHROMID(TM) CPS ELITE 418284 manufactured by Biomerieux Sa.

Event Text Entries

[29213716] Device not returned to manufacturer.
Patient Sequence No: 1, Text Type: N, H10


[29213717] On (b)(6) 2015, a customer from (b)(6) reported an incorrect identification of a bacteria associated with using chromid(tm) cps elite translucen ref. 418284, lot 1004156150 exp. 10-nov-2015. The customer observed pink isolates on cpse agar, which according to the instructions for use would indicate presence of escherichia coli. The customer performed an antibiogram test and obtained a suspected result for carbapenemase. Therefore, the customer sent the strain to the reference lab to confirm the carbapenemase producing indicator. The reference lab identified the strain as hafnia alvei bacteria and excluded the carbapenemase producing indicator. The customer performed additional identification tests from isolates on cpse agar and cos agar (lot not provided): identification of hafnia alvei bacteria was confirmed. The customer did not provide the confirmation results to the physician, the false identification for escherichia coli was reported to the physician.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3002769706-2015-00127
MDR Report Key5173190
Report SourceFOREIGN,HEALTH PROFESSIONAL,U
Date Received2015-10-24
Date of Report2015-10-01
Date Mfgr Received2015-10-01
Device Manufacturer Date2015-07-15
Date Added to Maude2015-10-24
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. RYAN LEMELLE
Manufacturer Street595 ANGLUM ROAD
Manufacturer CityHAZELWOOD MO 63042
Manufacturer CountryUS
Manufacturer Postal63042
Manufacturer Phone3147318582
Manufacturer G1BIOMERIEUX SA
Manufacturer StreetCHEMIN DE L ORME
Manufacturer CityMARCY L ETOILE, RHONE 69280
Manufacturer CountryFR
Manufacturer Postal Code69280
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCHROMID(TM) CPS ELITE
Generic NameCHROMID(TM) CPS ELITE
Product CodeJXA
Date Received2015-10-24
Catalog Number418284
Lot Number1004156150
Device Expiration Date2015-11-10
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerBIOMERIEUX SA
Manufacturer AddressCHEMIN DE L ORME MARCY L ETOILE, RHONE 69280 FR 69280


Patients

Patient NumberTreatmentOutcomeDate
10 2015-10-24