CHROMID(TM) CARBA AGAR 20 PLT 43861

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2015-10-26 for CHROMID(TM) CARBA AGAR 20 PLT 43861 manufactured by Biomerieux Sa.

Event Text Entries

[29974328] Device not returned to manufacturer.
Patient Sequence No: 1, Text Type: N, H10


[29974332] Summary: a customer in italy reported a selectivity issue with the chromid carba agar (carb) (b)(4), lot 1004183450 exp. 12oct2015. The customer reported 3 different patient rectal swab samples tested using chomid carba agar, lot 1004183450, were observed to have growth of e. Coli; for patient 1. Klebsiella, for patient 2. A mix of enterococcus and candida for patient 3. Customer reported that all the strains from these three patients were confirmed to not be carbapenemase producing enterobacteriaceae (cpe) (by molecular tests from an external lab) and therefore the strains should not have grown on the carb plates. No wrong results have been reported however, customer reported some results have been delayed 48h. No patient was harmed as consequence of the reported event.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3002769706-2015-00128
MDR Report Key5177172
Date Received2015-10-26
Date of Report2015-10-07
Date Mfgr Received2015-10-07
Device Manufacturer Date2015-07-28
Date Added to Maude2015-10-26
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. RYAN LEMELLE
Manufacturer Street595 ANGLUM ROAD
Manufacturer CityHAZELWOOD MO 63042
Manufacturer CountryUS
Manufacturer Postal63042
Manufacturer Phone3147318582
Manufacturer G1BIOMERIEUX SA
Manufacturer StreetCHEMIN DE L ORME
Manufacturer CityMARCY L ETOILE, RHONE 69280
Manufacturer CountryFR
Manufacturer Postal Code69280
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameCHROMID(TM) CARBA AGAR 20 PLT
Generic NameCHROMID(TM) CARBA AGAR 20 PLT
Product CodeJSI
Date Received2015-10-26
Catalog Number43861
Lot Number1004183450
Device Expiration Date2015-10-12
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerBIOMERIEUX SA
Manufacturer AddressCHEMIN DE L ORME MARCY L ETOILE, RHONE 69280 FR 69280