BLOOD GAS BLOOD COLLECTION CAPILLARY TUBE 04996974

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 01,05 report with the FDA on 2004-03-25 for BLOOD GAS BLOOD COLLECTION CAPILLARY TUBE 04996974 manufactured by Bayer Corp..

Event Text Entries

[19487899] Blood gas capillary tube shattered upon introduction to the rapidlab 865 blood gas analyzer. Specimen was from a pt with known hep c. Operator has been treated per hosp protocol. It was noted later that the operator was not using the recommended capillary tube and they may have been placing additional stress on the device with their technique.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1217157-2004-00004
MDR Report Key517969
Report Source01,05
Date Received2004-03-25
Date of Report2004-03-25
Date of Event2004-03-15
Date Mfgr Received2004-03-15
Date Added to Maude2004-03-30
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location0
Manufacturer ContactMARY GRAY
Manufacturer Street*
Manufacturer CityMEDFIEDL MA 02052
Manufacturer CountryUS
Manufacturer Postal02052
Manufacturer Phone5083593826
Manufacturer G1MODULOHM A/S
Manufacturer StreetVASEKAER 6 - 8
Manufacturer CityHERLEV 2730
Manufacturer CountryDA
Manufacturer Postal Code2730
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameBLOOD GAS BLOOD COLLECTION CAPILLARY TUBE
Generic NameLITHIUM HEPARINIZED CAPILLARY TUBES
Product CodeGIO
Date Received2004-03-25
Model NumberNA
Catalog Number04996974
Lot Number0707227
ID Number*
Device Expiration Date2005-07-01
Device AvailabilityY
Device Eval'ed by MfgrN
Implant FlagN
Date RemovedA
Device Sequence No1
Device Event Key507004
ManufacturerBAYER CORP.
Manufacturer Address* E. WALPOLE MA 02032 US
Baseline Brand NameBLOOD GAS BLOOD COLLECTION CAPILLAR TUBE
Baseline Generic NameLITHIUM HEPARINIZED CAPILLARY TUBES
Baseline Model NoNA
Baseline Catalog No04996974
Baseline ID*


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2004-03-25

© 2026 FDA.report