STRYKERVISION FLAT-PANEL MOUNTING ARM 0240095200

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2015-10-27 for STRYKERVISION FLAT-PANEL MOUNTING ARM 0240095200 manufactured by Stryker Endoscopy-san Jose.

Event Text Entries

[29985404] The device manufacture date is not known at this time. However, should it become available it will be provided in future reports. Additional information will be provided once the investigation has been completed.
Patient Sequence No: 1, Text Type: N, H10


[29985445] It was reported the mount arm on the stryker vision broke at the elbow.
Patient Sequence No: 1, Text Type: D, B5


[33590779] The product was not returned for investigation at stryker endoscopy, therefore the reported failure mode was not confirmed. The failure mode will be monitored for future reoccurrence.?? The device was not received for investigation at stryker endoscopy. It was evaluated at tekna. The evaluation results indicates: "we have evaluated this arm. It looks like what happened is that the pivot pins worked loose from one side of the arm, resulting in higher stress on the other side causing the part to break. We have had couple of complaints like this in the past and had traced it to a rare reaming operation at our supplier. Back in the 2004, early 2005 time period, if the operators had difficulty pressing in the pins, they would occasionally ream the holes a little larger. This allows the pin to come loose later. This operation is no longer performed at the supplier. Even in that condition, the arm lasted over ten years though. "
Patient Sequence No: 1, Text Type: N, H10


[33590780] It was reported the mount arm on the stryker vision broke at the elbow.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number0002936485-2015-00963
MDR Report Key5180611
Date Received2015-10-27
Date of Report2015-10-02
Date of Event2015-09-29
Date Mfgr Received2015-10-02
Date Added to Maude2015-10-27
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. KIMBERLY LYNCH
Manufacturer Street5900 OPTICAL COURT
Manufacturer CitySAN JOSE CA 95138
Manufacturer CountryUS
Manufacturer Postal95138
Manufacturer Phone4087542000
Manufacturer G1STRYKER ENDOSCOPY-SAN JOSE
Manufacturer Street5900 OPTICAL COURT
Manufacturer CitySAN JOSE CA 95138
Manufacturer CountryUS
Manufacturer Postal Code95138
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameSTRYKERVISION FLAT-PANEL MOUNTING ARM
Generic NameCART, EMERGENCY, CARDIOPULMONARY (EXCLUDING EQUIPMENT)
Product CodeBZN
Date Received2015-10-27
Catalog Number0240095200
Lot Number040330039
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerSTRYKER ENDOSCOPY-SAN JOSE
Manufacturer Address5900 OPTICAL COURT SAN JOSE CA 95138 US 95138