MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2015-10-27 for API? 20 A STRIPS 20300 manufactured by Biomerieux Sa.
[30041316]
Device not returned to manufacturer.
Patient Sequence No: 1, Text Type: N, H10
[30041317]
On (b)(6) 2015, a customer contacted biomerieux to report a mis-identification of bifidobacterium species when using api 20 a strips. The customer conducted additional testing with the vitek ms and identified the sample as parabacteroides distasonya. The results from the vitek ms were communicated to the treating physician; there was no delay in results or patient treatment.
Patient Sequence No: 1, Text Type: D, B5
[35135632]
Biom? Rieux internal investigation was conducted. The strain submitted by the customer was tested via the following methods: api? 20 a strips (two strips each from customer lot and a random lot). Vitek? 2 anc id test kit. Vitek? Ms. All test methods returned the expected result of parabacteroides distasonis. The discrepant result obtained by the customer could not be reproduced. The api 20 a strip is performing in accordance with specifications.
Patient Sequence No: 1, Text Type: N, H10
Report Number | 3002769706-2015-00130 |
MDR Report Key | 5180884 |
Date Received | 2015-10-27 |
Date of Report | 2015-10-08 |
Date Mfgr Received | 2015-10-08 |
Device Manufacturer Date | 2015-03-12 |
Date Added to Maude | 2015-10-27 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | MR. RYAN LEMELLE |
Manufacturer Street | 595 ANGLUM ROAD |
Manufacturer City | HAZELWOOD MO 63042 |
Manufacturer Country | US |
Manufacturer Postal | 63042 |
Manufacturer Phone | 3147318582 |
Manufacturer G1 | BIOMERIEUX SA |
Manufacturer Street | CHEMIN DE L ORME |
Manufacturer City | MARCY L ETOILE, RHONE 69280 |
Manufacturer Country | FR |
Manufacturer Postal Code | 69280 |
Single Use | 3 |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | API? 20 A STRIPS |
Generic Name | API? 20 A STRIPS |
Product Code | JSP |
Date Received | 2015-10-27 |
Catalog Number | 20300 |
Lot Number | 1003848900 |
Device Expiration Date | 2016-04-24 |
Operator | HEALTH PROFESSIONAL |
Device Availability | N |
Device Age | DA |
Device Eval'ed by Mfgr | N |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | BIOMERIEUX SA |
Manufacturer Address | CHEMIN DE L ORME MARCY L ETOILE, RHONE 69280 FR 69280 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2015-10-27 |