MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2015-10-27 for API? 20 A STRIPS 20300 manufactured by Biomerieux Sa.

Event Text Entries

[30041316] Device not returned to manufacturer.
Patient Sequence No: 1, Text Type: N, H10


[30041317] On (b)(6) 2015, a customer contacted biomerieux to report a mis-identification of bifidobacterium species when using api 20 a strips. The customer conducted additional testing with the vitek ms and identified the sample as parabacteroides distasonya. The results from the vitek ms were communicated to the treating physician; there was no delay in results or patient treatment.
Patient Sequence No: 1, Text Type: D, B5


[35135632] Biom? Rieux internal investigation was conducted. The strain submitted by the customer was tested via the following methods: api? 20 a strips (two strips each from customer lot and a random lot). Vitek? 2 anc id test kit. Vitek? Ms. All test methods returned the expected result of parabacteroides distasonis. The discrepant result obtained by the customer could not be reproduced. The api 20 a strip is performing in accordance with specifications.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3002769706-2015-00130
MDR Report Key5180884
Date Received2015-10-27
Date of Report2015-10-08
Date Mfgr Received2015-10-08
Device Manufacturer Date2015-03-12
Date Added to Maude2015-10-27
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. RYAN LEMELLE
Manufacturer Street595 ANGLUM ROAD
Manufacturer CityHAZELWOOD MO 63042
Manufacturer CountryUS
Manufacturer Postal63042
Manufacturer Phone3147318582
Manufacturer G1BIOMERIEUX SA
Manufacturer StreetCHEMIN DE L ORME
Manufacturer CityMARCY L ETOILE, RHONE 69280
Manufacturer CountryFR
Manufacturer Postal Code69280
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameAPI? 20 A STRIPS
Generic NameAPI? 20 A STRIPS
Product CodeJSP
Date Received2015-10-27
Catalog Number20300
Lot Number1003848900
Device Expiration Date2016-04-24
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerBIOMERIEUX SA