MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2015-10-28 for TC NEEDLE HOLDER X-DEL SERR200MM BM026R manufactured by Aesculap Ag.
[30219563]
Manufacturing site evaluation: on-going.
Patient Sequence No: 1, Text Type: N, H10
[30219564]
Device broke during a hip fracture. An x-ray was performed and found to be negative for retained piece. No patient injury; surgery delay 10-15 minutes.
Patient Sequence No: 1, Text Type: D, B5
[36597115]
(b)(4). Manufacturing site evaluation: one of the hard metal inlays of the received instrument was broken 2 mm from the tip. The solder is partially dissolved. Investigation was completed visually, using digital microscope keyence vhx-600. The breakage is user/maintenance related. The breakage appears to have resulted from a needle slipping off the needle holder. Aggressive maintenance methods can result in dissolving of the solder, causing the carbide metal to become brittle. Corrective/preventive action is not required.
Patient Sequence No: 1, Text Type: N, H10
Report Number | 2916714-2015-00989 |
MDR Report Key | 5183931 |
Date Received | 2015-10-28 |
Date of Report | 2016-01-21 |
Date Mfgr Received | 2015-09-28 |
Date Added to Maude | 2015-10-28 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | MS. NICOLE BROYLES |
Manufacturer Street | 615 LAMBERT POINTE DRIVE |
Manufacturer City | HAZELWOOD MO 63042 |
Manufacturer Country | US |
Manufacturer Postal | 63042 |
Manufacturer Phone | 3145515988 |
Manufacturer G1 | AESCULAP AG |
Manufacturer Street | PO BOX 40 |
Manufacturer City | TUTTLINGEN, 78501 |
Manufacturer Country | GM |
Manufacturer Postal Code | 78501 |
Single Use | 3 |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | TC NEEDLE HOLDER X-DEL SERR200MM |
Generic Name | NEEDLE HOLDER |
Product Code | HXK |
Date Received | 2015-10-28 |
Model Number | BM026R |
Catalog Number | BM026R |
Device Availability | * |
Device Age | DA |
Device Eval'ed by Mfgr | N |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | AESCULAP AG |
Manufacturer Address | PO BOX 40 TUTTLINGEN, 78501 GM 78501 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 1. Other | 2015-10-28 |