Event Text Entries

[30219563] Manufacturing site evaluation: on-going.
Patient Sequence No: 1, Text Type: N, H10


[30219564] Device broke during a hip fracture. An x-ray was performed and found to be negative for retained piece. No patient injury; surgery delay 10-15 minutes.
Patient Sequence No: 1, Text Type: D, B5


[36597115] (b)(4). Manufacturing site evaluation: one of the hard metal inlays of the received instrument was broken 2 mm from the tip. The solder is partially dissolved. Investigation was completed visually, using digital microscope keyence vhx-600. The breakage is user/maintenance related. The breakage appears to have resulted from a needle slipping off the needle holder. Aggressive maintenance methods can result in dissolving of the solder, causing the carbide metal to become brittle. Corrective/preventive action is not required.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2916714-2015-00989
MDR Report Key5183931
Date Received2015-10-28
Date of Report2016-01-21
Date Mfgr Received2015-09-28
Date Added to Maude2015-10-28
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. NICOLE BROYLES
Manufacturer Street615 LAMBERT POINTE DRIVE
Manufacturer CityHAZELWOOD MO 63042
Manufacturer CountryUS
Manufacturer Postal63042
Manufacturer Phone3145515988
Manufacturer G1AESCULAP AG
Manufacturer StreetPO BOX 40
Manufacturer CityTUTTLINGEN, 78501
Manufacturer CountryGM
Manufacturer Postal Code78501
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameTC NEEDLE HOLDER X-DEL SERR200MM
Generic NameNEEDLE HOLDER
Product CodeHXK
Date Received2015-10-28
Model NumberBM026R
Catalog NumberBM026R
Device Availability*
Device AgeDA
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerAESCULAP AG
Manufacturer AddressPO BOX 40 TUTTLINGEN, 78501 GM 78501


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2015-10-28

© 2026 FDA.report