[16792957] Operator used o2 under pressure to dry the inside of the cal tube. When the o2 tubing was placed into the tube, the exit hole of the device was occluded. This caused over pressure of the calibration tube. As a result, the end of the cal tube separated from cyclinder. This flew through the air striking the operator in the face, broke their glasses and knocked them to the floor. Cal tube is not designed for this type of cleaning procedure.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2023374-2004-00001
MDR Report Key519027
Report Source05
Date Received2004-04-02
Date of Report2004-04-01
Date of Event2004-03-26
Date Mfgr Received2004-03-26
Date Added to Maude2004-04-06
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location0
Manufacturer ContactGUY HARRIS
Manufacturer Street9150 COMMERCE CENTER CIRCLE #500
Manufacturer CityHIGHLANDS RANCH CO 80129
Manufacturer CountryUS
Manufacturer Postal80129
Manufacturer Phone8004684556
Manufacturer G1*
Manufacturer Street*
Manufacturer City*
Manufacturer Country*
Single Use3
Remedial ActionOT
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameSANDHILL PRESSURE CAL TUBE
Generic NamePART NUMBER S90-5119
Product CodeBXY
Date Received2004-04-02
Model NumberS90-5119
Catalog NumberNA
Lot NumberNA
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Implant FlagN
Date RemovedA
Device Sequence No1
Device Event Key508057
ManufacturerSANDHILL SCIENTIFIC
Manufacturer Address9150 COMMERCE CENTER CIR.,#500 HIGHLANDS RANCH CO 80129 US
Baseline Brand NameSANDHILL PRESSURE CAL TUBE
Baseline Generic NamePART NUMBER S90-5119
Baseline Model NoS90-5119
Baseline Catalog NoNA
Baseline IDNA