OSHER MALYUGIN MANIPULATOR MAL-0003

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2015-11-02 for OSHER MALYUGIN MANIPULATOR MAL-0003 manufactured by Microsurgical Technology Inc.

Event Text Entries

[30570128] The device evaluation showed that the manipulator experienced extreme forces that bent and racked the ball of the device prior to use. The damage most likely occured during maintenance.
Patient Sequence No: 1, Text Type: N, H10


[30570129] A piece of the osher/malyugin manipulator came off in the patient's eye. Piece was removed without issue and surgery was completed as planned.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3019924-2015-00033
MDR Report Key5195832
Date Received2015-11-02
Date of Report2015-10-30
Date Mfgr Received2015-10-02
Date Added to Maude2015-11-02
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. ROBERT MAY
Manufacturer Street8415 154TH AVE NE
Manufacturer CityREDMOND WA 98052
Manufacturer CountryUS
Manufacturer Postal98052
Manufacturer Phone4255560544
Manufacturer G1MICROSURGICAL TECHNOLOGY INC
Manufacturer Street8415 154TH AVE NE
Manufacturer CityREDMOND WA 98052
Manufacturer CountryUS
Manufacturer Postal Code98052
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameOSHER MALYUGIN MANIPULATOR
Generic NameOPHTHALMIC HOOK
Product CodeHNQ
Date Received2015-11-02
Returned To Mfg2015-10-13
Model NumberMAL-0003
Catalog NumberMAL-0003
OperatorPHYSICIAN
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerMICROSURGICAL TECHNOLOGY INC
Manufacturer Address8415 154TH AVE NE REDMOND WA 98052 US 98052


Patients

Patient NumberTreatmentOutcomeDate
10 2015-11-02