TDXFLX SYSTEM 04A24-96

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2015-11-04 for TDXFLX SYSTEM 04A24-96 manufactured by Abbott Manufacturing Inc.

Event Text Entries

[30617431] Further investigation of the customer issue included instrument maintenance, a review of the complaint text, a search for similar complaints and a review of labeling. The fse replaced the reagent display door with board (part number (b)(4)), which was worn out from normal use and verified the system to be functional again after replacement. No adverse trend was identified for the customer's issue. Labeling was reviewed and found to be adequate. Based on all available information and abbott diagnostics' complaint investigation, no product deficiency was identified.
Patient Sequence No: 1, Text Type: N, H10


[30617432] The customer observed visible smoke coming from the tdx analyzer. The field service engineer (fse) identified the source of the smoke as coming from the electronic area of the door. No injuries were reported.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1628664-2015-00273
MDR Report Key5201146
Date Received2015-11-04
Date of Report2015-11-04
Date Mfgr Received2015-10-13
Device Manufacturer Date1992-03-10
Date Added to Maude2015-11-04
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag0
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactNOEMI ROMERO-KONDOS, RN BSN
Manufacturer Street100 ABBOTT PARK ROAD DEPT. 09B9, LCCP1-3
Manufacturer CityABBOTT PARK IL 600643537
Manufacturer CountryUS
Manufacturer Postal600643537
Manufacturer Phone224667-512
Manufacturer G1ABBOTT MANUFACTURING INC
Manufacturer Street1921 HURD DRIVE
Manufacturer CityIRVING TX 75038
Manufacturer CountryUS
Manufacturer Postal Code75038
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameTDXFLX SYSTEM
Generic NameAUTOMATED FLUORESCENT IMMUNOASSAY ANALYZER
Product CodeJJQ
Date Received2015-11-04
Catalog Number04A24-96
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerABBOTT MANUFACTURING INC
Manufacturer Address1921 HURD DRIVE IRVING TX 75038 US 75038


Patients

Patient NumberTreatmentOutcomeDate
10 2015-11-04

© 2026 FDA.report