VIVIDIMAGE T 26" SURGICAL GRADE MONITOR

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2015-11-04 for VIVIDIMAGE T 26" SURGICAL GRADE MONITOR manufactured by Vts Medical Systems.

Event Text Entries

[30625140] Please note the user facility reported steris isomedix as the manufacturer of the monitor. The actual manufacturer of the monitor is vts medical systems. A steris field service technician arrived onsite, inspected the surgical monitor, and identified the scalar required replacement. The monitor subject of the reported event uses an external converter that routes incoming data through a fiber optic cable to a scalar. The scalar then detects the signal and converts the data into a resolution which is compatible to be displayed on the monitor. The technician replaced the scalar, tested the unit, and confirmed the system to be operating according to specification. The unit was returned to service and no additional issues have been reported. The surgical monitor is not under steris contract agreement for maintenance.
Patient Sequence No: 1, Text Type: N, H10


[30625141] The user facility reported via medwatch 4508440000-2015-8009 that during a patient procedure their vividimage t 26" surgical grade monitor did not operate properly displaying a green screen. The or staff attempted to resolve the image displayed on the screen however they were not able to do so. The procedure was successfully completed with a different surgical monitor.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1000404456-2015-00007
MDR Report Key5201253
Date Received2015-11-04
Date of Report2015-11-04
Date of Event2015-06-04
Date Mfgr Received2015-10-05
Date Added to Maude2015-11-04
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. KATHRYN CADORETTE
Manufacturer Street5960 HEISLEY ROAD
Manufacturer CityMENTOR OH 44060
Manufacturer CountryUS
Manufacturer Postal44060
Manufacturer Phone4403927231
Manufacturer G1VTS MEDICAL SYSTEMS
Manufacturer Street40 MELVILLE PARK RD
Manufacturer CityMELVILLE NY 11747
Manufacturer CountryUS
Manufacturer Postal Code11747
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameVIVIDIMAGE T 26" SURGICAL GRADE MONITOR
Generic NameSURGICAL MONITOR
Product CodeKQM
Date Received2015-11-04
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerVTS MEDICAL SYSTEMS
Manufacturer Address40 MELVILLE PARK RD MELVILLE NY 11747 US 11747


Patients

Patient NumberTreatmentOutcomeDate
10 2015-11-04

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.