SARNS 8000 CARDIOPLEGIA MONITOR 16414

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2015-11-05 for SARNS 8000 CARDIOPLEGIA MONITOR 16414 manufactured by Terumo Cardiovascular Systems Corp..

Event Text Entries

[30709635] (b)(4). The ccp stated that they used a manometer as a backup. The field service representative (fsr) could not duplicate the reported issue. The fsr checked the cpg monitor operation and the cpg pressure monitor worked properly. The unit operated to manufacturer specifications and was returned to clinical use.
Patient Sequence No: 1, Text Type: N, H10


[30709636] It was reported that during set-up of the device for a cardiopulmonary bypass procedure, one of the pressure channels on the cardioplegia (cpg) monitor was not reading. As a result, an alternate device was employed. The perfusionist (ccp) used a manometer. The surgical procedure was completed successfully. There was no delay, no blood loss, nor adverse consequences to the patient.
Patient Sequence No: 1, Text Type: D, B5


[34306205] (b)(4). If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1828100-2015-00948
MDR Report Key5204650
Date Received2015-11-05
Date of Report2015-12-23
Date of Event2015-10-13
Date Mfgr Received2015-12-23
Date Added to Maude2015-11-05
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. JAN WINDER
Manufacturer Street6200 JACKSON ROAD
Manufacturer CityANN ARBOR MI 48103
Manufacturer CountryUS
Manufacturer Postal48103
Manufacturer Phone7346634145
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameSARNS 8000 CARDIOPLEGIA MONITOR
Generic NameALARM, BLOOD-PRESSURE-SARNS 8000 CARDIOPLEGIA MONITOR
Product CodeDSJ
Date Received2015-11-05
Model Number16414
Catalog Number16414
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerTERUMO CARDIOVASCULAR SYSTEMS CORP.
Manufacturer Address6200 JACKSON ROAD ANN ARBOR MI 48103 US 48103


Patients

Patient NumberTreatmentOutcomeDate
10 2015-11-05

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.