ABVISER AUTOVALVE IAP MONITORING WITH TRANSDUCER, POLE MOUNT ABV321

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative report with the FDA on 2015-11-05 for ABVISER AUTOVALVE IAP MONITORING WITH TRANSDUCER, POLE MOUNT ABV321 manufactured by Wolfe Tory Medical Inc.

Event Text Entries

[30545213] Convatec is submitting this report as a result of activities related to convatec remediation protocol (b)(4). Any additional information received regarding this event after filing this report will be filed on a supplemental report (medwatch 3500a).
Patient Sequence No: 1, Text Type: N, H10


[30545214] It was reported urine retention with the abviser. Bladder pressure completed and within a few minutes they noted the 20ml of flush in the urimeter. She did not see any urine in the urimeter or bag. They did bladder pressures every 6 hours for approximately 3 days so approximately 11 infusions for a total of 220ml. They performed a bladder scan and it read 300ml. They removed the abviser autovalve; reconnected the urimeter tubing and 250ml of urine drained out. The first bladder pressure was performed at 10am and a cystogram performed around noon on the same day. Technician instilled dye for cystogram in the foley catheter port. Nurse reported autovalve remains inflated even after removing it.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1722554-2015-31257
MDR Report Key5205240
Report SourceCOMPANY REPRESENTATIVE
Date Received2015-11-05
Date of Report2014-02-27
Date Mfgr Received2014-02-27
Date Added to Maude2015-11-05
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactJEANETTE JOHNSON
Manufacturer Street211 AMERICAN AVE
Manufacturer CityGREENSBORO NC 27409
Manufacturer CountryUS
Manufacturer Postal27409
Manufacturer Phone3362973009
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameABVISER AUTOVALVE IAP MONITORING WITH TRANSDUCER, POLE MOUNT
Generic NameDEVICE, CYSTOMETRIC, HYDRAULIC
Product CodeFEN
Date Received2015-11-05
Model NumberABV321
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerWOLFE TORY MEDICAL INC
Manufacturer Address79 WEST 4500 SOUTH SUITE 18 SALT LAKE CITY UT 84107 US 84107


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 2015-11-05

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.