LEICA RM2235 1492235EUUL

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a other report with the FDA on 2015-11-09 for LEICA RM2235 1492235EUUL manufactured by Leica Biosystems Nussloch Gmbh.

Event Text Entries

[30833431] An investigation of the incident is currently ongoing and a follow up including the outcome of the investigation will be submitted.
Patient Sequence No: 1, Text Type: N, H10


[30833432] On (b)(6) 2015, leica biosystems received a complaint that a customer got an injury to his index finger and middle finger during the usage of the microtome. Medical treatment was necessary.
Patient Sequence No: 1, Text Type: D, B5


[33606242] The investigation revealed the following: the incident was user related. During the usage of the microtome the user grabbed to a tool which was stored next to the microtome device. Whilst this action the users jacket came to the knife protector of the device and pushed the blade guard downwards. While returning back to the work position the user grasped into the blade of the knifo holder with his left hand, causing lacerations of the index and middle fingers.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number8010478-2015-00007
MDR Report Key5212156
Report SourceOTHER
Date Received2015-11-09
Date of Report2015-10-13
Date of Event2015-10-12
Date Mfgr Received2015-10-13
Device Manufacturer Date2015-01-30
Date Added to Maude2015-11-09
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. ROBERT GROPP
Manufacturer StreetHEIDELBERGERSTR. 17-19
Manufacturer CityNUSSLOCH, 69226
Manufacturer CountryGM
Manufacturer Postal69226
Manufacturer Phone6224143
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameLEICA RM2235
Generic NameMICROTOME
Product CodeIDO
Date Received2015-11-09
Model Number1492235EUUL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerLEICA BIOSYSTEMS NUSSLOCH GMBH
Manufacturer AddressHEIDELBERGERSTR. 17-19, NUSSLOCH, 69226 GM 69226


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2015-11-09

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.