URISYS 1100 03617556001

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a user facility report with the FDA on 2015-11-10 for URISYS 1100 03617556001 manufactured by Roche Diagnostics.

Event Text Entries

[30873923] It is not known if the initial reporter has or intends to report the event to fda.
Patient Sequence No: 1, Text Type: N, H10


[30873924] Reporter stated that back-to-back testing was performed on the urisys 1100 which, for test 1, yielded a protein result of 100 mg/dl and a glucose result of 250 mg/dl, while test 2 showed values of negative for protein and normal for glucose. No adverse event was reported. The manufacturer requested the return of the suspect product for evaluation.
Patient Sequence No: 1, Text Type: D, B5


[35048328] No investigation was done on the suspect device as the urisys 1100 device was not returned. The customer did return the chemstrip 10 md urine test strips. All results using the returned strips were in accordance to the testing plan. There were no false negative results at the urisys 1100 device observed.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1823260-2015-04501
MDR Report Key5213872
Report SourceUSER FACILITY
Date Received2015-11-10
Date of Report2016-01-04
Date of Event2015-10-06
Date Mfgr Received2015-10-14
Date Added to Maude2015-11-10
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactNA MICHAEL LESLIE
Manufacturer Street9115 HAGUE ROAD NA
Manufacturer CityINDIANAPOLIS IN 46250
Manufacturer CountryUS
Manufacturer Postal46250
Manufacturer Phone3175214343
Manufacturer G1ROCHE DIAGNOSTICS GMBH
Manufacturer StreetSANDHOFERSTRASSE 116 NA
Manufacturer CityMANNHEIM (BADEN-WURTTEMBERG) 68305
Manufacturer CountryGM
Manufacturer Postal Code68305
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameURISYS 1100
Generic NameAUTOMATED URINE ANALYZER
Product CodeKHE
Date Received2015-11-10
Model NumberNA
Catalog Number03617556001
Lot NumberNA
ID NumberNA
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerROCHE DIAGNOSTICS
Manufacturer Address9115 HAGUE ROAD NA INDIANAPOLIS IN 46250 US 46250


Patients

Patient NumberTreatmentOutcomeDate
10 2015-11-10