COC THC MOP AMP MAMP BZO BAR MTD MDMA OX DOA-1127-041-19-ONW02A

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional report with the FDA on 2015-11-12 for COC THC MOP AMP MAMP BZO BAR MTD MDMA OX DOA-1127-041-19-ONW02A manufactured by Alere San Diego, Inc..

Event Text Entries

[31019327] Investigation conclusion: customer's observation was not replicated in-house with retention product. Retention products were tested with 29 samples of drug-free donor urine; all mtd, coc results were negative at read time. No false positive were obtained. Mfg batch record review did not uncover any abnormalities. Root cause could not be determined from the information provided. Based on the information available, there is no indication of a product deficiency. No corrective action is required at this time.
Patient Sequence No: 1, Text Type: N, H10


[31019328] Distributor claiming their client was having false positive results for coc and mtd. No test lines developed in the test region for each drug, all control lines developed. Client unable to provide lab results but claimed they did confirmation testing and it was negative. No images or returns are available for investigation. No additional or patient information provided. No adverse patient sequela reported.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2027969-2015-00925
MDR Report Key5220507
Report SourceHEALTH PROFESSIONAL
Date Received2015-11-12
Date of Report2015-10-15
Date Mfgr Received2015-10-15
Date Added to Maude2015-11-12
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactYA-LING KING
Manufacturer Street9975 SUMMERS RIDGE ROAD
Manufacturer CitySAN DIEGO CA 92121
Manufacturer CountryUS
Manufacturer Postal92121
Manufacturer Phone8588052084
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCOC THC MOP AMP MAMP BZO BAR MTD MDMA OX
Generic NameDRUG OF ABUSE TEST
Product CodeDKZ
Date Received2015-11-12
Model NumberDOA-1127-041-19-ONW02A
Lot NumberDOA5030217
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerALERE SAN DIEGO, INC.
Manufacturer Address9975 SUMMERS RIDGE ROAD SAN DIEGO CA 92121 US 92121


Patients

Patient NumberTreatmentOutcomeDate
10 2015-11-12

© 2026 FDA.report