FRESENIUS DIALYSIS DELIVERY SYSTEM 2008K NA

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05 report with the FDA on 2004-04-21 for FRESENIUS DIALYSIS DELIVERY SYSTEM 2008K NA manufactured by Fresenius Medical Care-north America.

Event Text Entries

[310162] A dialysis facility reported that the venous line was found disconnected from the pt's subclavian catheter during a hemodialysis treatment. There was no machine alarm. Blood was found on the floor but the amount of blood lost could not be estimated. Venous pressure (vp) at 7:32 a. M. (prior to the incident) was 136. At 7:46 a. M. , when the disconnection was discovered, the vp reading was 77. The venous alarm width selected is unknown. The pt was taken to the hosp and expired a few days later. Contact does not have any info as to the cause of death or if an autopsy was performed.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2937457-2004-00007
MDR Report Key522074
Report Source05
Date Received2004-04-21
Date of Report2004-03-22
Date of Event2004-03-12
Date Facility Aware2004-03-12
Report Date2004-03-22
Date Reported to Mfgr2004-03-22
Date Mfgr Received2004-03-22
Device Manufacturer Date2001-01-01
Date Added to Maude2004-04-27
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactELVY DIZON, BSN, RN, CNN
Manufacturer Street2637 SHADELANDS DR
Manufacturer CityWALNUT CREEK CA 94598
Manufacturer CountryUS
Manufacturer Postal94598
Manufacturer Phone8002272572
Manufacturer G1*
Manufacturer Street*
Manufacturer City*
Manufacturer Country*
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameFRESENIUS DIALYSIS DELIVERY SYSTEM
Generic NameHEMODIALYSIS MACHINE
Product CodeFIL
Date Received2004-04-21
Model Number2008K
Catalog NumberNA
Lot NumberNA
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Age3 YR
Device Eval'ed by MfgrY
Implant FlagN
Date RemovedA
Device Sequence No1
Device Event Key511204
ManufacturerFRESENIUS MEDICAL CARE-NORTH AMERICA
Manufacturer Address* * *
Baseline Brand NameFRESENIUS DIALYSIS DELIVERY SYSTEM
Baseline Generic NameHEMODIALYSIS MACHINE
Baseline Model No2008K
Baseline Catalog NoF002-40
Baseline IDNA
Baseline Device FamilyNA
Baseline Shelf Life [Months]NA
Baseline PMA FlagN
Baseline 510K PMNY
Premarket NotificationK994267
Baseline PreamendmentN
Baseline TransitionalN
510k ExemptN


Patients

Patient NumberTreatmentOutcomeDate
101. Death 2004-04-21

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.