BLOM-SINGER ADVANTAGE INDWELLING VOICE PROSTHESIS AD2012-NS

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a consumer report with the FDA on 2015-11-16 for BLOM-SINGER ADVANTAGE INDWELLING VOICE PROSTHESIS AD2012-NS manufactured by Freudenberg Medical, Llc Dba Helix Medical, Llc.

Event Text Entries

[31320007]
Patient Sequence No: 1, Text Type: N, H10


[31320008] On (b)(4) 2015: the patient sent an email to freudenberg medical stating they had 2 emergency room visits to find where their indwelling voice prosthesis had gone. They initially went to the emergency room and they took x-rays and they could not see it. "they assumed it went down the esophagus even though" the patient stated they were "breathing heavily. " the patient returned to the emergency room. The doctor listened to their lungs and "could tell something was wrong by the right lung. " a cat scan was done which showed "where it was located. " the pulmonologist took the patient to the operating room, gave the patient anesthesia and removed it. The removed device was given to the patient who noted that one of the flanges was missing. On (b)(4) 2015: an slp from freudenberg medical spoke to the patient and the following information was provided: patient "is in good health - no adverse health issues as a result of the incident. " product was placed in (b)(6) 2014. Timeline/events per patient report: approximately 1 week before the prosthesis reportedly dislodged, they "experienced some gastric issues" which they "now retrospectively think was related to the esophageal flange falling off and passing through" their digestive system. During that same week, the patient noticed that the voice prosthesis appeared to be pistoning. The "valve dislodged" and the patient placed a patient changeable voice prosthesis. The patient had a chest x-ray which was negative. The patient felt short of breath and called slps who told them to go for further evaluation as it was possible that the aspirated prosthesis did not show up on the x-ray. The patient had a ct of the lung which showed the presence of the prosthesis in their airway. The patient underwent a bronchoscopy and a pulmonologist removed the prosthesis. Subsequently, the patient was in good health. On 9/4/2015: device was received at freudenberg medical with a note from the patient which stated: "enclosed is defective advantage indwelling. Removed with bronchoscope by pulmonologist... On (b)(6) 2015. "
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2025182-2015-00005
MDR Report Key5227133
Report SourceCONSUMER
Date Received2015-11-16
Date of Report2015-11-16
Date of Event2015-08-08
Date Mfgr Received2015-08-10
Date Added to Maude2015-11-16
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. BELINDA JACKSON
Manufacturer Street1110 MARK AVENUE
Manufacturer CityCARPINTERIA CA 93013
Manufacturer CountryUS
Manufacturer Postal93013
Manufacturer Phone8055765308
Manufacturer G1FREUDENBERG MEDICAL, LLC DBA HELIX MEDICAL, LLC
Manufacturer Street1110 MARK AVENUE
Manufacturer CityCARPINTERIA 93013
Manufacturer CountryUS
Manufacturer Postal Code93013
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameBLOM-SINGER ADVANTAGE INDWELLING VOICE PROSTHESIS
Generic NameVOICE PROSTHESIS
Product CodeEWL
Date Received2015-11-16
Returned To Mfg2015-09-04
Model NumberAD2012-NS
Catalog NumberAD2012-NS
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerFREUDENBERG MEDICAL, LLC DBA HELIX MEDICAL, LLC
Manufacturer Address1110 MARK AVENUE CARPINTERIA CA 93013 US 93013


Patients

Patient NumberTreatmentOutcomeDate
101. Hospitalization; 2. Required No Informationntervention 2015-11-16

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.