PYXIS ANESTHESIA SYSTEM (PAS) 4000

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative report with the FDA on 2015-11-18 for PYXIS ANESTHESIA SYSTEM (PAS) 4000 manufactured by Carefusion.

Event Text Entries

[31571778] (b)(4). Additional data / failure investigation. Field service technician on site reported the customer spilled iv solution on the device which ingressed into the ups and caused a short out.
Patient Sequence No: 1, Text Type: N, H10


[31571779] Customer reports seeing a small amount of smoke and hearing a popping sound from the ups on the pyxis anesthesia system.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2016493-2015-00003
MDR Report Key5234374
Report SourceCOMPANY REPRESENTATIVE
Date Received2015-11-18
Date of Report2015-11-04
Date of Event2015-11-04
Date Mfgr Received2015-11-04
Device Manufacturer Date2014-03-06
Date Added to Maude2015-11-18
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactCARL IVERSON
Manufacturer Street10020 PACIFIC MESA BLVD.
Manufacturer CitySAN DIEGO CA 92121
Manufacturer CountryUS
Manufacturer Postal92121
Manufacturer Phone8586175207
Manufacturer G1CAREFUSION
Manufacturer Street10020 PACIFIC MESA BLVD.
Manufacturer CitySAN DIEGO CA 92121
Manufacturer CountryUS
Manufacturer Postal Code92121
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NamePYXIS ANESTHESIA SYSTEM (PAS)
Generic NameAUTOMATED DISPENSING CABINET (ADC'S)
Product CodeBRY
Date Received2015-11-18
Model Number4000
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerCAREFUSION
Manufacturer Address10020 PACIFIC MESA BLVD. SAN DIEGO CA 92121 US 92121


Patients

Patient NumberTreatmentOutcomeDate
10 2015-11-18

© 2026 FDA.report