INTREPID AUTOSERT IOL HANDPIECE 8065751755

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a other report with the FDA on 2015-11-24 for INTREPID AUTOSERT IOL HANDPIECE 8065751755 manufactured by Alcon Research, Ltd. - Irvine Technology Center.

Event Text Entries

[32021754] Investigation including root cause analysis is in progress. A supplemental mdr will be filed as necessary in accordance with 21 cfr 803. 56 when additional reportable information becomes available. (b)(4)
Patient Sequence No: 1, Text Type: N, H10


[32021755] A customer reported that an intraocular lens (iol) handpiece injector cracked the iol cartridge during a procedure. The procedure was completed. The patient experienced a larger wound opening and wound leakage. Additional information has been requested but none has been received.
Patient Sequence No: 1, Text Type: D, B5


[37285308] The company representative reported? During iol implantation the injectors are cracking the cartridge.? Additional information has been requested but has not been received. The company service representative examined the injectors and determined that all three found, would require replacement, one of which was determined to have had a bent tip. The root cause cannot be determined conclusively. (b)(4).
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2028159-2015-09610
MDR Report Key5246415
Report SourceOTHER
Date Received2015-11-24
Date of Report2016-02-02
Date of Event2015-11-02
Date Mfgr Received2016-02-01
Date Added to Maude2015-11-24
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. EDDIE DARTON, MD, JD
Manufacturer Street6201 SOUTH FREEWAY MAIL STOP AB2-6
Manufacturer CityFORT WORTH TX 76134
Manufacturer CountryUS
Manufacturer Postal76134
Manufacturer Phone8175686660
Manufacturer G1ALCON RESEARCH, LTD. - IRVINE TECHNOLOGY CENTER
Manufacturer Street15800 ALTON PARKWAY
Manufacturer CityIRVINE CA 92618
Manufacturer CountryUS
Manufacturer Postal Code92618
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameINTREPID AUTOSERT IOL HANDPIECE
Generic NameAPPARATUS, CAUTERY, RADIOFREQUENCY, AC-POWERED
Product CodeHQR
Date Received2015-11-24
Model NumberNA
Catalog Number8065751755
Lot NumberASKU
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerALCON RESEARCH, LTD. - IRVINE TECHNOLOGY CENTER
Manufacturer Address15800 ALTON PARKWAY IRVINE CA 92618 US 92618


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2015-11-24