[32100827] A customer in (b)(6) reported their thinprep 5000 processor with autoloader etched a wrong vial id onto a slide. The error was caught by the customer at the time and no patients were affected. Hologic's field service engineer (fse) was unable to confirm the error. This is considered a reportable event since the thinprep 5000 processor did not perform as intended.
Patient Sequence No: 1, Text Type: D, B5


[38888776] After reassessing this incident it has been determined that the customer was using an unsupported label barcode symbology; therefore this is not a reportable event in the us as it was previously reported (1222780-2015-00214). As per the current operator manual (man-01783-001 rev 008) the only supported symbologies are code 128, interleaved 2 of 5, code 39 and code 93; the customer was using codabar symbology.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1222780-2015-00214
MDR Report Key5246561
Date Received2015-11-24
Date of Report2015-11-23
Date Mfgr Received2015-11-06
Date Added to Maude2015-11-24
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactEVA MAXWELL
Manufacturer Street250 CAMPUS DRIVE
Manufacturer CityMARLBOROUGH MA 01752
Manufacturer CountryUS
Manufacturer Postal01752
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameTHINPREP 5000 PROCESSOR
Generic NameTHINPREP 5000 PROCESSOR
Product CodeMKQ
Date Received2015-11-24
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerHOLOGIC, INC
Manufacturer Address250 CAMPUS DRIVE MARLBOROUGH MA 01752 US 01752


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2015-11-24