BD KIESTRA? READA COMPACT 447206

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2015-11-25 for BD KIESTRA? READA COMPACT 447206 manufactured by Bd Kiestra Lab Automation.

Event Text Entries

[32320235] The bd kiestra? Reada compact is the incubation and imaging module designed for use with the total laboratory automation (tla) and work cell automation (wca) lines of bd kiestra products. On 11/20/2015 bd quality confirmed during the investigation that a functionality setting in the bd kiestra? Reada compact system which automatically adjusts the timing between images of the dishes to account for any system downtime has been determined to contribute to the reported event. The customer? S settings were changed to prevent this event from reoccurring. Bd has determined through the investigation that no us customer site is impacted by this functionality setting on the bd kiestra? Reada compact system. No device returned.
Patient Sequence No: 1, Text Type: N, H10


[32320236] The customer is receiving an error pop up window while processing samples using the bd kiestra reada compact. The error itself can be resolved simply by restarting the software. However, if the error goes unresolved the system may begin skipping incubation steps. This caused the final images to be taken prior to the end of their originally programmed incubation periods.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3010141591-2015-00003
MDR Report Key5249108
Date Received2015-11-25
Date of Report2015-11-24
Date of Event2015-06-25
Date Mfgr Received2015-10-15
Date Added to Maude2015-11-25
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactBDX CAROL NIETO
Manufacturer Street7 LOVETON CIRCLE
Manufacturer CitySPARKS MD 21152
Manufacturer CountryUS
Manufacturer Postal21152
Manufacturer Phone4103164000
Manufacturer G1BD KIESTRA LAB AUTOMATION
Manufacturer Street6 MARCONILAAN
Manufacturer CityDRACHTEN
Manufacturer CountryUS
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameBD KIESTRA? READA COMPACT
Generic NameMICROBIOLOGICAL INCUBATOR
Product CodeJTQ
Date Received2015-11-25
Catalog Number447206
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerBD KIESTRA LAB AUTOMATION
Manufacturer Address6 MARCONILAAN DRACHTEN US


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2015-11-25

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.