LIGASURE BLUNT TIP OPEN SEALER/DIVIDER LF1520

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2015-11-30 for LIGASURE BLUNT TIP OPEN SEALER/DIVIDER LF1520 manufactured by Covidien Lp.

Event Text Entries

[32407416] (b)(4). Date of initial report: (b)(6) 2015. To date the incident sample has not been received for evaluation. If the sample is received or if additional information pertinent to the incident is obtained a follow-up report will be submitted.
Patient Sequence No: 1, Text Type: N, H10


[32407417] The customer reported that during a hysterectomy, the device jaws could not be re-opened. A new instrument of the same type was opened and used to complete the case. There was no patient injury or tissue damage.
Patient Sequence No: 1, Text Type: D, B5


[38297188] (b)(4). The reported condition was confirmed. The jaws were locked closed when received and the trigger was jammed. Excessive eschar was found between the jaws. The ski guide was caught on the internal track due to lateral force placed on the latch causing it to jump the a-track. The latch would not release to open the jaws. The device was disassembled and the investigation discovered damage to the ski tabs where the ski caught on the a-track. This occurs when the user applies excessive force to the latch. Manufacturing non-conformances were reviewed. No entries pertinent to the customer's report were noted.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1717344-2015-00855
MDR Report Key5255071
Date Received2015-11-30
Date of Report2015-11-24
Date of Event2015-11-18
Date Mfgr Received2015-12-31
Device Manufacturer Date2015-03-04
Date Added to Maude2015-11-30
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS SHARON MURPHY
Manufacturer Street5920 LONGBOW DRIVE
Manufacturer CityBOULDER CO 80301
Manufacturer CountryUS
Manufacturer Postal80301
Manufacturer Phone2034925267
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameLIGASURE BLUNT TIP OPEN SEALER/DIVIDER
Generic NameLIGASURE VESSEL SEALING SYSTEM
Product CodeBWA
Date Received2015-11-30
Returned To Mfg2015-11-09
Model NumberLF1520
Catalog NumberLF1520
Lot Number50160238X
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerCOVIDIEN LP
Manufacturer Address5920 LONGBOW DRIVE BOULDER CO 80301 US 80301


Patients

Patient NumberTreatmentOutcomeDate
10 2015-11-30